Honeywell ST 700 User Manual page 75

Smartline series
Hide thumbs Also See for ST 700:
Table of Contents

Advertisement

7.5.3.2 Correct Input at the Lower Range Value (LRV)
1. After the LRV and URV have been entered, as described in the previous Correct Input at
Zero procedure, click the Correct LRV button on the CALIBRATION display. (See Step 4
in the previous procedure to open the CALIBRATION display.)
2. Click the Correct LRV button. This message appears:
3. Adjust the PV input pressure to the exact value of the LRV entered in the DE CONFIGURE
display.
4. Click the OK button; this message appears:
5. Observe the input pressure at the applied value; when it is stable, click the OK button.
6. When the Transmitter has completed the LRV correction, this message appears:
7. Click OK to acknowledge.
Revision 4.0
ST 700 Series HART/DE Option User's Manual
Page 61

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents