Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

DR/2400
Method 8194
Scope and Application:
* Adapted from Standard Methods for the Examination of Water and Wastewater.
Tips and Techniques
• Distillation is required for determining total selenium. See the Distillation on page 4 at the end of the procedure. Use the distillate
in step 2.
• Acetone (Cat. No. 14429-49) is a suitable solvent for removing toluene from glassware after results are measured.
• Toluene (F005) solutions are regulated as hazardous waste by the Federal RCRA. Do not pour these materials down the drain.
Water saturated with toluene, toluene solutions, and the cotton plug used in the delivery tube of the separatory funnel should be
collected for disposal with laboratory solvent wastes. See Section 4 on page 55 for more information on proper disposal of these
materials.
• Wipe the outside of sample cells before each insertion into the instrument cell holder. Use a damp towel followed by a dry one to
remove fingerprints or other marks.
• If there are visible water bubbles on the bottom of the cell, decant the top portion into a clean, dry 10-mL cell prior to reading the
sample.
Hach Programs
1.
Touch
.
Hach Programs
Select program
.

640 Selenium

Touch
.
Start
Selenium_None_Other_DAB_Eng_Ody.fm
For water and wastewater
B B
S
2.
Measure 100 mL of
deionized water into a
500-mL Erlenmeyer flask.
Label the flask "blank".
Measure 100 mL of
sample into a 500-mL
Erlenmeyer flask. Label
the flask "sample".
Diaminobenzidine Method*
B
S
3.
Add a 0.2-g spoonful
®
of TitraVer
Hardness
Reagent to each flask.
Swirl to mix.
Selenium
(0.01 to 1.00 mg/L)
Method 8194
B
S
4.
Add a 0.05-g spoonful
of diaminobenzidene
tetrahydrochloride to
each flask. Swirl to mix.
Selenium
Page 1 of 8

Advertisement

Table of Contents
loading

Table of Contents