Iron, Total - Hach DR2400 Manual

Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

DR/2400
(
Method 8008
Powder Pillows or AccuVac
Scope and Application:
total iron; USEPA approved for reporting wastewater analysis**
* Adapted from Standard Methods for the Examination of Water and Wastewater
** Federal Register, June 27, 1980; 45 (126:43459)
Tips and Techniques
• Digestion is required for determining total iron for EPA reporting purposes. See the Hach Water Analysis Handbook for digestion
procedures.
• For more accurate results, determine a reagent blank value for each new lot of reagent. Follow the procedure using deionized
water in place of the sample. Subtract the reagent blank value from the final results or perform a reagent blank adjust
instrument manual for more information on Running a Reagent Blank.
• After adding reagent, an orange color will form if iron is present.
• Accuracy is not affected by undissolved powder.
Powder Pillows
Hach Programs
1.
Touch
.
Hach Programs
Select program
.
265 Iron, FerroVer
Touch
.
Start
IronTot_AVPP_Other_FVR_Eng_Ody.fm
®
Ampuls
For water, wastewater, and seawater; digestion is required for determining
2.
Fill a clean, round
sample cell with 10 mL of
sample.
3.
Add the contents of
one FerroVer Iron
Reagent Powder Pillow
to the sample cell (the
prepared sample). Swirl
to mix.

Iron, Total

®
FerroVer
Method*
(0.02 to 3.00 mg/L)
.
See the
Method 8008
4.
Touch the timer icon.
Touch
.
OK
A three-minute reaction
period will begin.
(Allow samples that
contain rust to react for at
least 5 minutes.)
Iron, Total
Page 1 of 6