Hide thumbs Also See for DR2400:
Table of Contents

Advertisement

DR/2400
Method 8031
Powder Pillows or AccuVac
Scope and Application:
treated water, estuary water, and seawater
* Adapted from Palin, A.T., Inst. Water Eng., 21 (6), 537-547 (1967).
Tips and Techniques
• Analyze samples immediately. Do not preserve for later analysis.
• For more accurate results, determine a reagent blank value for each new lot of reagent. Follow the procedure using deionized
water in place of the sample. Subtract the reagent blank value from the final results or perform a reagent blank adjust
instrument manual for more information on Running a Reagent Blank.
• If the sample temporarily turns yellow after reagent addition, dilute a fresh sample. Repeat the test. A slight loss of iodine may
occur due to the dilution. Apply the appropriate dilution factor. See Section 2.7 Sample Dilution on page 21.
• Wipe the outside of sample cells before each insertion into the instrument cell holder. Use a damp towel followed by a dry one to
remove fingerprints or other marks.
Powder Pillows
Hach Programs
1.
Touch
.
Hach Programs
Select program
.

240 Iodine

Touch
.
Start
Iodine_AVPP_Other_DPD_Eng_Ody.fm
®
Ampuls
For testing dissolved iodine residual used as disinfectant in process water,
2.
Fill a cell with 10 mL
of sample.
3.
Add the contents of
one DPD Total Chlorine
Powder Pillow to the
sample cell (this is the
prepared sample).
Swirl to mix.
A pink color will develop
if iodine is present.
Iodine
DPD Method*
(0.07 to 7.00 mg/L)
.
See the
Method 8031
4.
Touch the timer icon.
Touch
.
OK
A three-minute reaction
period will begin.
Iodine
Page 1 of 4

Advertisement

Table of Contents
loading

Table of Contents