Section 1 Specifications - Hach Lange CL17 User Manual

Chlorine analyzer
Hide thumbs Also See for Lange CL17:
Table of Contents

Advertisement

Section 1
General
Display
Enclosure
Instrument description
Mounting
Instrument shipping weight
Warranty
Certification
Sample Requirements
Sample flow rate to sample conditioning
Inlet pressure to instrument
Inlet pressure to sampling conditioning
Sample temperature range
Inlet fitting at instrument
Drain fitting
Sample conditioning
Regent/Standard Requirements
Maximum reagent usage
Reagent containers
Reagent containment
Electrical
Power requirements
Power connection
Installation category
Polution degree
Protection class
Alarm relay outputs
Alarm connection
Recorder output
Recorder output connections
Specifications
Specifications are subject to change without notice.
LCD, 3½-digit measurement readout and six-character alphanumeric scrolling
text line.
IP62-rated with the gasketed door latched
31.4 cm (W) x 45.4 cm (H) x 17.9 (D) (12.4 x 17.88 x 7.06 in.)
Wall mount
10.49 kg (23 lb)
Hach Company warrants its products to the original purchaser against any
defects that are due to faulty material or workmanship for a period of one year
from the date of shipment unless otherwise noted in the product manual.
CE approved. Listed to UL and CSA safety standards by ETL.
200 to 500 mL/min
1 to 5 psig; 1.5 psig is optimum. Exceeding 5 psig can cause sample
tubing failure unless sample conditioning is used.
1.5 to 75 psig (with sample tube level with the bottom of the instrument—see
Figure 7 on page
5 to 40 °C (41 to 104 °F)
¼-inch OD polyethylene tubing with quick-disconnect fitting
½-inch hose barb
Use provided sample conditioning
One-half liter per month (each of the two reagents)
High-density polyethylene (2) ½-liter bottles
Reagent bottles are contained inside the analyzer enclosure and are vented
externally.
100–115/230 VAC (selection switch inside instrument); 90 VA, 50/60 Hz, 2.5 Amp
fuse
Connection made by three wire barrier terminal block through a ½-inch conduit
hole in the case. Wire range: 18 AWG.
II
2
I
Two unpowered SPDT relays each rated at 5 A resistive, 240 V ac maximum.
Can function as sample set point alarms (high or low) or as a system warning
indicator or a system alarm indicator. Use either high voltage (greater than 30 V
RMS and 42.2 V PEAK or 60 V dc) or low voltage (less than 30 V RMS and 42.2
V PEAK or 60 V dc). Do not use a combination of high and low voltage.
Connection made by a removable three wire plug through a ½-inch conduit hole
in the case. Wire range: 12–18 AWG.
One isolated recorder output, 4–20 mA (can be adjusted to 0–20 mA).
Recommended load impedance 3.6 to 500 ohms
Connection made by a removable three wire plug through a ½-inch conduit hole
in the case. Wire range: 12–22 AWG.
16.)