Maintenance

6.2.2 Clean the instrument enclosure

6.2.3 Clean the colorimeter

*
Do not use a plastic swab when cleaning the colorimeter with sulfuric acid. The the acid will dissolve the plastic.
42
With the enclosure securely latched, use a soft cloth and a mild detergent to wipe the
outside of the enclosure. Do not allow moisture to enter the enclosure.
The colorimeter measuring cell may collect sediment or develop a film growth on the
inside walls. Cleaning with an acid solution and a cotton swab on a monthly basis is
recommended. Depending on sample conditions, it may be necessary to clean the cell
more often.
CAUTION
Chemical exposure hazard. The chemicals used in this procedure may be
hazardous if inappropriately handled or accidentally misused. Please read all
warnings on the reagent labels. Protective eye wear is always recommended when
contact with chemicals is possible.
Colorimeter cell cleaning procedure:
1. Press the
key, then press the down
MENU
appears. Press
ENTER
2. Press the down
ARROW
3. Refer to
Figure 22
and locate the colorimeter.
4. Remove the rubber plug on the top of the colorimeter assembly.
5. Remove the magnetic stir bar using a bent steel paper clip as a retrieval tool. Refer
to
Figure
22.
6. Wait until
flashes on the status line of the display, then fill the colorimeter with
CLEAN
19.2 N Sulfuric Acid Standard Solution (203832).
Note: Lower normality sulfuric acid will NOT be sufficient for cleaning the colorimeter.
7. Allow the sulfuric acid to stand in the colorimeter for 15 minutes.
8. Insert a wooden or paper cotton-tipped swab
down with a gentle scrubbing action, cleaning the interior surfaces of the colorimeter
cell. Clean the stir bar with the cotton swab
9. Install the stir bar.
10. Replace the rubber plug on the top of the colorimeter assembly and make sure the
cover is on the viewing port of the colorimeter.
11. Press the
key to immediately return to normal operation; the instrument will
EXIT
return to normal operation automatically after 60 minutes.
ARROW
.
key to scroll to the
CLEAN
*
into the opening and move it up and
(Figure
key until the
menu
MAINT
option. Press
.
ENTER
23).