A method to remove power from the relays locally must
be available in case of an emergency or for servicing the
instrument.
3.8.3

Voltage selection for alternate voltage operation

Figure 11 Customer connections to analyzer
Figure 12 Power connections
Note: The voltage selector switch must be set properly for the line voltage to be used. An improper
setting can result in serious damage to the instrument when power is applied. Refer to
The instrument is switched for 115-volt operation when it leaves the factory. To convert
the instrument for 230-volt operation, slide the AC line conversion switch (shown in
Figure
13) to the 230V position. The fuses used in this product are 5 mm x 20 mm. The
fuses used are approved for both North American and European use; they do not need to
be changed when voltage is converted.
3
AC power: 100-115/230 VAC, 50/60 Hz, 90 VA fused at
2.50A
2
Since the ON/OFF switch can be accessed without the
use of a tool an external switch is not required for
instrument power.
Installation
Figure
13.
21