Specifications - Hach APA 6000 Operation Manual

High range hardness process analyzer
Table of Contents

Advertisement

Specifications

Performance
Sample Requirements
Page 4
Specifications
Specifications are subject to change without notice.
See instrument-specific specifications in the APA 6000™ Process Analyzer
Manual, Cat. No. 62005-18.
Range:10 to 1000 mg/L as CaCO
Note: If sample is outside the specified instrument range, a sample out error
may occur.
Accuracy: ±5% of reading or ±2 mg/L as CaCO
Repeatability: ±5% of reading or ±2 mg/L as CaCO
Response Time: For single channel instruments, less than 17 minutes for
90% response to step change at sample inlet
Cycle Time: 8.2 minutes (average)
Sample Temperature Range: 5 to 50 °C (41 to 122 °F)
Operating Temperature: 5 to 50 °C (41 to 122 °F)
Relative Humidity: 5 to 95% relative humidity, non-condensing
Sample Pressure: 0.5 to 30.0 psig (0.03 to 2.04 bar)
Sample Temperature: 5 to 50 °C (41 to 122 °F)
Sample Flow: 100 to 2000 mL/min. maximum
Sample Inputs: Up to two sample streams
3
, whichever is greater
3
, whichever is greater
3
6200018Specs.fm