Guide to Operation
18
Figure 9 is a sample N-20P extended-mode printout.
(Note: This session was ended prematurely at 1 minute
57 seconds.)
NELLCOR N-20P
PULSE OXIMETER
VERSION 0.0.0
Patient
Identification
Area
Date, Time-
-{
PATIENT
•*foj
Indicator of 2%-
Sp02 Decrease
-12-31-99
23:59
SAT BPM
00:30 97% 70
01:00 99% 66
*01:12 97% 75
01:30 99% 70
-*01:48 97% 69.
Time from Start
Vj§i/
of Session, Sp02,
Pulse Rate
\^H/
Total-
Monitoring
Time
CASE SUMMARY:
2:
MIN:
97%
MAX:
99%
MEAN:
98%
SE RATE
MIN:
66
MAX:
75
MEAN:
70
TIME:
01:57
-Statistical
Summary
Figure 9: Extended-Mode Printout
On some printouts, an asterisk may appear to identify a 1%
decrease in Sp02; this is not the case. For example, if Sp02 is
98% at 30 seconds, increases to 99% at 37 seconds, and
decreases to 97% at 44 seconds, the printout would show:
00:30
SAT 98%
*00:44
SAT 97%
In this case, the asterisk appears to mark a 1% decrease, but
there was actually a 2% drop in Sp02 at 44 seconds (from 99%
to 97%).
Note:
Patient identification cannot be typed in.