15B MAX/17B MAX
Calibration Manual
Replace Batteries and Fuses
To prevent false readings, which could lead to possible electric shock or personal
injury, replace the batteries as soon as the battery indicator (
To prevent damage or injury, install ONLY replacement fuses with the specified
amperage, voltage, and interrupt ratings.
Disconnect test leads before you open the case or the battery door.
Dispose of old devices in a professional and environmentally appropriate manner.
Delete personal data on the Product before disposal.
Remove batteries that are not integrated into the electrical system before disposal.
Dispose of batteries separately.
To replace the batteries or fuses, see
B1
Performance Tests
Use the performance tests to make sure the Device Under Test (DUT) operates properly and to
make sure the DUT is accurate. If the DUT fails any part of the performance test, repair and/or
calibration adjustment are required. See
to meet the range indicated, see
4
XW Warning
Figure
2.
Figure 2. Replace Batteries and Fuses
Temperature
Contact
Fluke.
F2
F1
Connections. If the DUT continues to fail