Performance Test Procedures

The performance tests given in this section includes the Repeatability
Test. Perform each step in the order given, using the corresponding test
equipment.
Make sure that all optical connections of the test setup given in the
procedure are dry and clean. DO NOT USE INDEX MATCHING OIL. Make
sure that all optical connectors are undamaged.
For cleaning, use the cleaning instructions given in "Cleaning Information"
on page 35.
Make sure that all optical cables of the test setup are fixed to the table so
that they cannot move during measurements.
Movement of the fibers during the test procedures and the quality of
optical connectors affect the result of power measurements.
The environmental conditions (temperature and relative humidity) must
remain constant during the test.