Completion

Save/Load Defaults

Save Defaults
Load Defaults
3-32
Installation: Completion
Save/Load Defaults is used to back up/restore system defaults for archive and
disaster recovery. The defaults that are backed up and restored through the following
processes include the IP address, Computer Name, Tower name and receiver
allocation.
1.
Log in to the Webmin application according to
21.
2.
Click the Configuration tab.
3.
Click the Save/Load Defaults link under the Configuration menu along the left
side of the window.
4.
Click Backup.
5.
Click the ApexPro Backup Files link ending with the .tar extension.
6.
Click Save.
7.
Navigate to where you want to save the configuration file and click Save.
8.
Click Close to close the Download Complete window.
:$51,1*
LOSS OF MONITORING — The ApexPro Telemetry Server
software must be stopped before performing this procedure.
Notify the site's biomedical department so patient
communications can be established with an alternative central
station before stopping the ApexPro Telemetry Server software
(the Bed Duplicator application can be used to temporarily
transfer patients to an alternate ApexPro tower. See
Bed Transfer"
on page 4-4). Shut down of the ApexPro Telemetry
ApexPro Telemetry Server
2001989-178
"Webmin Logons"
on page 2-
"Temporary
Revision C