Chassis Fan - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Chassis Fan

6-38
Parts Lists, Drawings and Replacement: Disassembly/Assembly
1.
Complete
"Field Replacement Preparation"
2.
Remove the chassis cover according to
3.
Disconnect the cabinet fan power connectors from the receptacles along the left
edge of the SNMP connector board. Note the position and orientation of the
connectors for reassembly.
4.
Using a phillips screwdriver, remove the 2 screws securing the cabinet fan
assembly at the front of the chassis. Remove the fan assembly.
5.
Remove the fan from the fan assembly.
6.
Reverse the steps in this procedure to install the CPU fan.
7.
If applicable, perform the
ApexPro Telemetry Server
2001989-178
on page 6-5.
"Chassis Cover"
"Electrical Safety Tests"
on page 4-10.
on page 6-10.
022A
101A
Revision C