Download Print this page

Advertisement

CONTENTS
■TU-6100 specifications-------------------------------1
■Instrument descriptions------------------------------2
■Display descriptions----------------------------------3
■Key descriptions--------------------------------------4
■Instructions-------------------------------------------5
■RS-232 transfer protocol-----------------------------8
■Precautions------------------------------------------11

Advertisement

loading
Need help?

Need help?

Do you have a question about the TU-6100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Rixen TU-6100

  • Page 1: Table Of Contents

    CONTENTS ■TU-6100 specifications-------------------------------1 ■Instrument descriptions------------------------------2 ■Display descriptions----------------------------------3 ■Key descriptions--------------------------------------4 ■Instructions-------------------------------------------5 ■RS-232 transfer protocol-----------------------------8 ■Precautions------------------------------------------11...
  • Page 2: Tu-6100 Specifications

    Dimensions 150 x 75 x 28 mm,Approx.320g (battery included) Input connection Four wire connector waterproof socket TU-6100 High accuracy thermometer LP-12U Immersion temperature probe(Class A)-ψ 2.8 ×L150mm LP-32U Needle temperature probe (Class A)-ψ 2.8 × L120mm Standard TU-655C Pouch case...
  • Page 3: Instrument Descriptions

    ■Instrument descriptions ○ Large LCD Display ○ Battery cover ○ Power switch ○ AC Adaptor Input ○ Setting and ℃/℉ switching ○ USB Output port ○ Function key ○ RS-232 Output connector ○ Setting and RTD changing ○ T1 Probe input connector ○...
  • Page 4: Display Descriptions

    ■Display descriptions Display Descriptions Battery Power Symbol Manual shut down LED Back light on Buzzer on Under Calibrating Maximum Value HOLD Data Hold Minimum Value Average Value T1 probe T2 probe PT-100Ω/PT-1KΩ Temperature Probe Type Hi.A High point Alarm Lo.A Low point Alarm Area Digit A.B.C.
  • Page 5: Key Descriptions

    ■Key descriptions ※The RTD type and the ℃/℉ is changeable only under the testing mode. ※After inserted the probe will run internal calibration in few seconds. Key Function Descriptions LCD display Press once for power on or off. (If the instrument is left without Power any button operation, it will automatically turn off after five On/Off...
  • Page 6: Instructions

    ■Instructions A. Testing Mode: Press to switch display conditions. Probe condition, Probe type, warning data condition, Normal conditions Power indication T1 Real-time T1 Temp. data, (Area A) Temperature display T1 Hi.A data (Area B), T1 Lo.A data(Area C) T2 Real-time T2 Temp.
  • Page 7 B. Readings hold mode Under the testing mode, press to enter hold mode, and press to return to the testing mode. Press key to switch the display condition. The data will be deleted when the mode or system are changed. Condition Display content T1-HOLD display...
  • Page 8 C. Setting Mode Under the testing mode, press over 2 seconds to enter the setting mode. Press for over 2 seconds to switch to the next setting mode. Push key to increase the value, Push key to reduce the value Push key to change digit When all settings are done, it will return to the testing mode.
  • Page 9: Rs-232 Transfer Protocol

    ■RS-232 transfer protocol RS-232 the transmission end is one way transfer via the three wires (touch ground) to input and output. Using the original transmission line or cable under 10m is recommended. Baud rate: 57600, 19200, 9600, 4800 Transfer status: / 8 / N / 1 Transfer content: (8BIT) Read Holding Registers by Function 03H A.
  • Page 10 Slave Byte Data(1) Data(1) Data (1) Function Data(1) Hi address count Middle Last Data(2) Data (2) Data(2) Data (2) Last Middle 0x00013883=80003,Real value = 80003/100,Real value = 800.03 0x00013CAB=81067,Real value = 81067/100,Real value = 810.67 ☆℃ and ℉ set is data of MOD “bit 05” at address 0x000A C .Request Data Frame Ex: Read the data from address 0x0008 (Read 10-byte of data from address 0x0008) Slave...
  • Page 11 **Example** Set mode 1. Temperature unit=℃ 2. RTD=PT100 3. T1 no ERROR 4. T2 no ERROR 5. Power74~50% Address Data name Data size 0000h T1 Real-Time Data(Hold) 4-Byte 0001h T1 Max 4-Byte 0002h T1 Min 4-Byte 0003h T1 Avg 4-Byte 0004h T2 Real-Time Data(Hold) 4-Byte...
  • Page 12: Precautions

    1. This instrument has a waterproof function; please do not use it in a high temperature Environment or with corrosive materials to avoid leakage or damage. 2. To avoid the problems that might occur when connecting to the computer, the Rixen USB computer interface cable and windows software (Model No. TU-USB-C) is recommended.