Resources for OmniPod System users are available
from the Customer Care tab on our Web site:
MyOmniPod.com
Among these resources are an online copy of this
User Guide and the My OmniPod System Information
form, which you will also find at the end of this
guide.
If the PDM cannot communicate with the Pod to accept or shut
off a hazard alarm, you can shut it off manually.
1. Remove the Pod from your body if you have not already
done so.
2. Press the end of a paper clip or similar item into the manual
alarm shut-off port on the top side of the Pod. (See Chapter
11, Communication Failures, Figure 11-17.)
Errors, Advisories, and Hazard Alarms
Alarm
Empty reservoir.
Insulin delivery stopped.
Change pod now.
Occlusion detected.
Insulin delivery stopped.
Change pod now.
Pod expired.
Insulin delivery stopped.
Change pod now.
Auto-off.
Remove pod now.
Source Action
Pod
Press OK to acknow-
and
ledge.
PDM
Change Pod.
Check blood glucose.
Pod
Press OK to acknow-
ledge.
Change Pod.
Check blood glucose.
Pod
Press OK to acknow-
ledge.
Change Pod.
Check blood glucose.
Pod
Press OK to acknow-
and
ledge.
PDM
Change Pod.
Check blood glucose.
10
127