3.2 Battery safety and usage

Batteries should be stored in an ambient temperature of -4°F (–20°C) to
67.9°F (+20°C). Lithium batteries will self-discharge during storage, so
they should be fully charged before storage. Self-discharge may affect long
term battery performance. Furthermore, it is recommended that the battery
should be charged in a set time, as below;
Ambient temperature of:
-4°F (–20°C) to 67.9°F (+20°C), once every 6 months
67.9°F (+20°C) to 113°F (+45°C), once every 3 months
113°F (+45°C) to 149°F (+60°C), once every month
Battery should have greater than 50% charge prior to recharge.
The ambient temperature for charging should be 32°F (0°C) to 104°F
(+40°C). If under 32°F (0°C), the battery capacity will be diminished; if
over 104°F (+40°C), the battery may overheat, resulting in permanent
damage.
WARNING:
Never disassemble, incinerate or puncture battery
Never short circuit the battery
Keep battery dry
Keep out of reach of children
Always dispose of battery in accordance with local, state and federal
regulations

3.3 Quick access

Capture an image

3.3.1
Witch on the black body, set the blackbody temperature at 37℃
After installing the battery, press and hold the power key (≥3 seconds)
until the display comes on. After approximately 50 seconds, the power
on routine will finish and the camera will start its measurements.
Press button "C" to switch the display image into thermal mode
Remove the lens cap, point the camera at the blackbody, and adjust
16