Led Indications, Mains Connections And Potential Differences - Gossen MetraWatt PROFITEST MBASE MTECH Operating Instructions Manual

Test instruments for iec 60364.6 / din vde 0100
Hide thumbs Also See for PROFITEST MBASE MTECH:
Table of Contents

Advertisement

18.1

LED Indications, Mains Connections and Potential Differences

Status Test
Plug
LED Indications
NETZ/
lights up
X
green
MAINS
NETZ/
blinks
MAINS
green
NETZ/
X
blinks red
MAINS
NETZ/
lights up
red
MAINS
NETZ/
lights up
yellow
MAINS
lights up
U
/R
X
L
L
red
lights up
FI/RCD
X
red
Mains Connection Test — Single-Phase System
LCD Connection Pictograph
?
is
displayed
?
?
PE
is
displayed
L
N
PE
is
displayed
L
N
PE
is
displayed
L
N
PE
is
x
displayed
L
N
PE
is
x
displayed
L
N
PE
is
x
displayed
L
N
PE
is
displayed
L
N
PE
is
displayed
x
L
N
PE
is
displayed
L
N
Mains Connection Test — 3-Phase System
LCD Connection Pictograph
is
displayed
is
displayed
is
displayed
GMC-I Messtechnik GmbH
Meas.
Selector Switch
Adapter
Position
I
N / I
F
Z
/ Z
/ R
L-N
L-PE
E
I
N / I
F
X
Z
/ Z
/ R
L-N
L-PE
E
I
N / I
F
X
Z
/ Z
/ R
L-N
L-PE
E
X
R
/ R
ISO
LO
I
/ I
N
F
X
Z
/ Z
/ R
L-N
L-PE
E
I
N
X
R
/ R
ISO
LO
X
I
/ I
N
F
All except for U
All except for U
All except for U
All except for U
All except for U
All except for U
All except for U
All except for U
All except for U
All except for U
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
Function / Meaning
Correct connection, measurement enabled
N conductor not connected,
measurement enabled
1) no line voltage or
2) PE interrupted
Interference voltage detected, measurement disabled
L and N are connected with the phase conductors.
– Contact voltage U
or U
IN
– Safety shut-down has occurred
– Limit value exceeded or fallen short of for R
The RCCB was not tripped, or was tripped too late during the
tripping test.
No connection detection
Connection OK
L and N reversed, neutral conductor charged with phase voltage
No mains connection
Neutral conductor interrupted
Protective conductor PE interrupted, neutral conductor N and/or phase
conductor L charged with phase voltage
Phase conductor L interrupted, neutral conductor N charged with phase
voltage
Phase conductor L and protective conductor PE reversed
Phase conductor L and protective conductor PE reversed
Neutral conductor interrupted (only with probe)
L and N are connected with the phase conductors.
Clockwise rotation
Counter-clockwise rotation
Short between L1 and L2
> 25 V or > 50 V
I
/ R
function
INS
LO
51

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROFITEST MBASE MTECH and is the answer not in the manual?

Table of Contents