5.5
Help Function
The following information can be displayed for each switch
position and basic function after it has been selected with the rotary
selector switch:
• Wiring diagram
• Measuring range
• Nominal range of use and Measuring Uncertainty
• Nominal value
Ð Press the HELP key in order to query online help:
Ð If several pages of help are available for the respective
measuring function, the HELP key must be pressed repeatedly.
Ð Press the ESC key in order to exit online help.
5.6
Setting Parameters or Limit Values using RCD Measurement as an Example
4
1 Access the submenu for setting the desired parameter.
2 Select the parameter with the orscroll key.
3 Go to the settings menu for the selected parameter with the scroll key.
4 Adjust the value with the orscroll key.
5 Acknowledge the setting value with the key. This value is transferred to
the settings menu.
6 The setting value is not permanently accepted for the
respective measurement until
returned to the main menu. You can return to the main menu by pressing
3
ESC instead of
without accepting the newly selected value.
GMC-I Messtechnik GmbH
1
2
4
4
5
5
3
is pressed, after which the display is
2
3
2
3
6
Parameter Blocking
The individual selected parameters are subjected to a plausibility
test before they are accepted by the measuring window.
If the selected parameter does not make sense
in combination with the other parameters that
have been preset, it is not accepted and an
error window appears.
The parameter previously set is retained in the
memory.
Remedy: Select another parameter.