3. To Disinfect: Using fresh wipes, make sure that all outside
surfaces of the lancing device remain wet for 2 minutes.
Super Sani-Cloths may be purchased at the following places:
• Amazon.com,
• Officedepot.com or visit your local store,
• Walmart.com
4. Let lancing device air dry thoroughly
before using to test. Replace End Cap.
Gently pull back and release Arming
Barrel and press the Trigger Button.
A click will be heard if the lancing
device is functioning properly.
5. Wash hands thoroughly again after
handling the lancing device.
Stop using the lancing device and call Customer Care for
assistance at 1-800-803-6025, available 24 hours a day, 7
days a week, if:
• Markings on lancing device are coming off,
• Trigger Button hard to push,
• End Cap does not go back on,
• Arming Barrel does not click when gently pulled back.
Note: Other disinfectants have not been tested. The effect of other disinfec-
tants used interchangeably has not been tested with the meter. Use of
disinfectants other than Super Sani Cloth Wipes may damage meter.
Note: Super Sani Cloth Wipes have been tested on the lancing device for
a total of 1,095 cleaning and disinfecting cycles, which is equal to
cleaning and disinfecting the lancing device once per day for a 3 year
period. Lancing device may be cleaned and disinfected up to 7 cycles
per week for 3 years.