Introduction

The EMS Precision Ultrasound Balance has been developed in response to
the increasing demand for equipment capable of testing and calibrating
ultrasound therapy units.
The balance is based on the principle of measuring the radiation force
produced on a suitable target by an ultrasound transducer.
The force is measured using a precision digital electronic balance
programmed for direct read-out in watts. The unit is portable with carrying
case.
Due to the fact that the unit is factory set, if any problems arise with the
unit it is highly recommended that the unit is returned for service and re-
calibration.
Environmental Considerations
For best results the Precision Ultrasound Balance should only be used in an
environment free from excessive air currents or vibration. Do NOT install
the balance near open windows or doors causing draughts or rapid
temperature fluctuations, near heating or air-conditioning vents, near
vibrating or rotating machinery, or on an uneven surface.