The Electronic Balance may be cleaned by wiping over with a clean
damp cloth. The use of abrasive materials and cleaning solvents should
be avoided.
Inspect the mains adaptor periodically for signs of damage, especially cable
insulation.
The water used in the container must be changed at least weekly.
Periodically remove the rubber absorbers from the container and clean
them using warm soapy water. Rinse thoroughly after cleaning.
The calibration of the Ultrasound Balance may be checked periodically
using the 10 W weight provided with the balance:-
Zero the balance.
Place the 10 W weight on the target frame assembly beneath the
surface of the water.
Check that the Balance reads 10 W ± 0.5 W.
If the reading is not within the above range it is recommended that
the Ultrasound Balance is returned to EMS for calibration.
Calibration of the Ultrasound Balance requires specialised test equipment
and the balance must be returned to EMS Physio Ltd.