Specifications - Pintek PA-655 Instruction Manual

Pa-655 50 amp ac/dc current probe dc?500khz. 5ma?50a

Advertisement

Specifications:

These characteristics apply to an adjusted PA-655 AC/DC Current Probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with the
temperature at 10℃~30℃ and the humidity at 0~80.
Table 2: Electrical Characteristics
Current Ranges
DC Accuracy, typical
Gain versus frequency, typical
Maximum Working Current
Maximum Working Voltage
Maximum Float Voltage
Frequency Range
Rise Time
Battery Type and Life, typical
DC signal linearity, typical
Phase shift, typical
 
 
 
20/200 mV/A
±3% ±30 mA at 200 mV/A
(30 mA to 5 A peak range)
±4% ±300 mA at 20 mV/A
(300 mA to 20 A peak range)
±15% max at 200 mV/A
(20 A peak to 50 A peak range)
See Figure 6
See Table 3
See Table 3
See Table 3
DC to 500 kHz (-3 dB)
μS 
0.7
9V NEDA 1604A, IEC 6LR61
13 hours minimum (1 each)
See Figure 8
See Figure 9
13
www.tmatlantic.com
(Typ.)