Provider setup menu (cont.)
Alarm settings
Apnea
Patient disconnect
Low min vent
Low Vte
System settings
Backlight
Language**
hPa/cmH
O
2
Setup parameter display
Humidifier
Reset therapy hours*
Reset blower hours
Provider mode
2
*for firmware version 2.11 or higher
**for firmware version 2.24 or higher
Off, 10, 20, 30 sec
Off, 15, 60 sec
Off (0) to 99 lpm in 1.0 increments
On, Off
On, Off
English, Italian, Spanish, German, French, Danish, Dutch,
Norwegian, Swedish, Finnish, icons
hPa, cmH
O
2
Leak, RR, MinVent, Vte
1
0 to 5
Yes, No
Yes, No
On, Off
1. Setup parameter display
You can select which measured parameters will display on the
Monitor screen. Choose from Leak, RR (Respiratory Rate),
Min vent, or Vte. See the Measured Parameter section of the
User Manual for more information.
2. Provider mode
You can choose "on" or "off" to enable or disable provider
mode. Selecting "on" will unlock the device and give access
to all ventilation settings in Setup menu. Selecting "no" will
lock the device so the patient does not have access to his
ventilation settings.