Specifications - Verity Medical NeuroTrac Sports Operator's Manual

Hide thumbs Also See for NeuroTrac Sports:
Table of Contents

Advertisement

Specifications

STIM
1.
Dual channel: individually isolated circuits.
2.
Amplitude: 0-90 mA; indication only; actual mA will tend to be less
than indicated due to electrode impedance.
3.
Type: Constant Current.
4.
Waveform: Asymmetrical, rectangular bi-phasic with zero DC
current.
5.
Selectable pulse width: 50µS – 450µS [2% accuracy].
6.
Pulse Rate selection: in the continuous mode 2 – 100 Hz
[2% accuracy].
7.
Time duration of the treatment selectable: 1 minute to 90 minutes.
8.
Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2
volts the battery symbol will flash on/off once every second.
9.
Open Electrode Detect: If an open circuit is detected at the
output of channel A or B the output current will be reset at zero.
10.
Ramp up time 0.3 - 9.9 seconds.
11.
If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn
on.
12.
Physical dimensions: 108 x 62 x 23 mm.
13.
Weight: 0.07KG without battery, 0.1KG with battery.
14.
Environmental conditions for storage & transport:
-10 to +50 degrees Centigrade
0-90% Humidity.
NeuroTrac™ Sports Operation Manual