Advertisement

Quick Links

User guide
English

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Power Stayion II and is the answer not in the manual?

Questions and answers

Summary of Contents for ResMed Power Stayion II

  • Page 1 User guide English...
  • Page 2 Power Station II Battery pack / Batterie / Batería / Bateria Illustrations only, parts may differ / Illustrations uniquement, les pièces sont susceptibles d’être différentes / Solo ilustraciones, las piezas pueden diferir / apenas ilustrações; os itens reais podem ser diferentes Mute on/off Level Check Level...
  • Page 3: Power Station

    Not all devices are available in all regions. VPAP III ST-A with QuickNav and VS III are not available in the U.S. • For information on battery run times refer to the Battery/Compatibility List on www.resmed.com. Remarques : • Certains de ces appareils ne sont pas disponibles dans tous les pays. La VPAP III ST-A QuickNav et le VS III ne sont pas disponibles aux États-Unis.
  • Page 4: Intended Use

    English Intended use The ResMed Power Station II (RPS II) is an external lithium-ion battery that provides power when mains power is unavailable. Please read the entire guide before using the RPS II. Refer to the device user guides for intended patients, uses and contraindications associated with CPAP , bilevel and ventilation therapy.
  • Page 5 Setting up CAUTION • Ensure the RPS II and the device are turned off before setting up. • When the RPS II is plugged to the mains power supply, ensure all cables are properly connected. Charging the battery Refer to illustration C. Attach the power supply unit adapter to the DC plug and connect into the battery DC input/output port.
  • Page 6 3. 3a. When charging two batteries, attach the power supply unit adapter to the DC plug and connect the adapter to the battery coupler. 3b. When powering a device, connect the DC output cable of the device to the end of the battery coupler.
  • Page 7: Cleaning And Maintenance

    60% of the time from its original condition. A full charge in an older battery will not last as long as in a new battery. ResMed recommends that the RPS II be tested for battery duration regularly. As with all electrical equipment, if any irregularity becomes apparent, you should exercise caution and contact an authorised ResMed service representative.
  • Page 8: Troubleshooting

    Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the battery. Problem/Possible cause Solution Device is not running Power connections are disrupted.
  • Page 9: Technical Specifications

    (EMC) according to IEC60601- 1-2, for residential, commercial, and light industry environments. Information regarding the electromagnetic emissions and immunity of these ResMed devices can be found on www.resmed.com, on the Products page under Service and Support. Click on the PDF file for your language.
  • Page 10: General Warnings And Cautions

    If you need information on these disposal systems, please contact your local waste administration. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment.
  • Page 11: Limited Warranty

    • Only use a supplied ResMed 90W AC power supply unit for charging the RPS II. • Explosion hazard—do not use in the vicinity of flammable anesthetics. • The RPS II system is not intended for use by persons (including children) with reduced physical, sensory or mental capabilities, or lack of experience and knowledge, unless they have been given supervision or instruction concerning use of the RPS II system by a person responsible for their safety.
  • Page 12 • Portable diagnostic/screening devices This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
  • Page 14 Designs pending. AirSense, AirCurve, H5i, S9, Stellar and VPAP are trademarks of ResMed Ltd. S9, Stellar and VPAP are registered in U.S. Patent and Trademark Office. VSIII is a trademark of ResMed Paris SAS. Velcro is a registered trademark of Velcro Industries B.V.

This manual is also suitable for:

Rps ii

Table of Contents