Safety and Related Information
Guidance and Manufacturer's Declaration for Electromagnetic Immunity
The Kodak DryView 8900 Laser Imager is intended for use in the electromagnetic environment specified below. The
customer or user of the laser imager should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electrostatic
discharge (ESD):
• EN61000-4-2
• IEC 61000-4-2
Electrical fast
transient/burst:
lines
• EN61000-4-4
lines
• IEC 61000-4-4
Surge:
• EN61000-4-5
• IEC 61000-4-5
Voltage dips, short
<5% U * (>95% dip in
interruptions and
U *) for 0.5 cycle
voltage variations
40% U * (60% dip in
on power supply
U *) for 5 cycles
lines:
70% U * (30% dip in
• EN61000-4-11
U *) for 25 cycles
• IEC 61000-4-11
<5% U * (>95% dip in
U *) for 5 sec.
Power frequency
3 A/m
(50/60 Hz)
magnetic field:
• EN61000-4-8
• IEC 61000-4-8
NOTE: * U is the a.c. mains voltage prior to application of the test level.
1-14
6 kV contact
8 kV air
2 kV for power supply
1 kV for input/output
1 kV differential mode
2 kV common mode
6 kV contact
8 kV air
2 kV for power supply
lines
1 kV for input/output
lines
1 kV differential mode
2 kV common mode
<5% U * (>95% dip in
U *) for 0.5 cycle
40% U * (60% dip in
U *) for 5 cycles
70% U * (30% dip in
U *) for 25 cycles
<5% U * (>95% dip in
U *) for 5 sec.
3 A/m
5E6155
Electromagnetic Environment -
Guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Kodak
DryView 8900 Laser Imager requires
continued operation during power
mains interruptions, it is recommended
that the Kodak DryView 8900 Laser
Imager be powered from an
uninterruptible power supply or a
battery.
Mains power quality should be that of a
typical commercial or hospital
environment.