Stopping a Measurement
To stop an ongoing measurement operation, press the
causes the Base Unit to deflate the cuff and then go into standby.
button also causes the Base Unit to deflate the cuff
Pressing the
and turn off.
Recording Blood Pressure Readings
You can make copies of the "Blood Pressure Record" at the back of this
manual and use them to keep track of changes in your blood pressure
and pulse measurements.

ERROR MESSAGES

When any of the errors described below occurs, the Base Unit deflates the
cuff and then waits for further button operation or data communication with
the Watch.
Message
Cause
• The Base Unit cannot
Err 01
detect a suitable pulse.
Err 02
setting is too low.
• Noise due to body
movement.
• The cuff is not connected
Err 03
to the Base Unit properly,
or a hole in the cuff.
• The cuff is too loose on the
arm.
value is too low.
Err 04
• Noise due to body
movement.
• The cuff is too loose on the
arm.
• Increase in cuff pressure
.
Err 44
0
due to body movement.
(kPa)
Err 330
(mmHg)
The appearance of the low battery indicator shown below causes the
Base Unit to cancel any ongoing measurement or communication
operation and wait for further button operation.
Indicator
Cause
• Battery power is low.

TROUBLESHOOTING

Check the points noted below whenever you experience operational
problems with the Base Unit.
Problem
• Is battery power low?
Nothing happens when
you press the
→ Replace the batteries with a full set of
button.
new ones.
• Are batteries loaded correctly?
→ Check to make sure that batteries are
loaded correctly.
• Is the cuff's plug connected correctly to the
Inflation is performed, but
Base Unit?
the cuff does not inflate.
→ Check the plug for proper connection.
• Is there a hole in the cuff?
→ Contact the store or dealer where you
purchased the Base Unit about replacing
the cuff.
If the actions recommended above do not correct your problem, the
Base Unit itself may be malfunctioning. In this case, contact the store or
dealer where you purchased the Base Unit.
button. This
Action
• Do not move or speak
while measurement is in
progress.
• Make sure the air tube is
not under your arm.
• Check the air tube connec-
tion to the Base Unit.
• Tighten the cuff so there is
no space between your
arm and the cuff.
• If there is air leaking from
the cuff because it is old or
it has a hole in it, replace
the cuff with a new one.
• Do not move or speak
while measurement is in
progress.
• Make sure the air tube is
not under your arm.
• Tighten the cuff so there is
no space between your
arm and the cuff.
• Take care so no pressure
is applied to the cuff while
measurement is in
progress.
Action
Replace the Base Unit
batteries.
What to Check and What to Do

PRECAUTIONS DURING USE

• Do not leave the Base Unit in areas exposed to direct sunlight or near
a heater.
• Do not expose the Base Unit to severe vibration or impact.
• To clean the exterior of the Base Unit, wipe it with a dry cloth or with a
cloth moistened with a weak solution of water and a mild neutral
detergent. Wring all excess liquid from the cloth before wiping. Never
use thinner, benzene, or other volatile agents for cleaning. Doing so
can discolor or damage the case.
• Do not submerse the cuff in water or launder it.
• Be sure to remove the batteries from the Base Unit if you do not plan
to use it for a long time.
CASIO COMPUTER CO., LTD. assumes no responsibility for any
loss, or any claims by third parties that may arise through the
use of the Base Unit.