# Press the START button.
• This starts inflation of the cuff and displays the pressure value.
• During measurement, the
indicate that the pulse is being detected. The
flashing and remains on the display as long as the pulse rate is
displayed after measurement is complete.
$ After measurement is complete, the displays alternates between the
systolic and diastolic blood pressure values, and the pulse rate.
Systolic Blood Pressure
% Press the ON/OFF button to turn off the Base Unit.
• An Auto Power Off feature automatically turns off the Base Unit if
you leave it on for about three minutes.
Note
• Avoid taking too many consecutive measurements with the Base Unit.
Pressure from the Base Unit cuff can cut off circulation in your arm
and cause incorrect readings. Remove the cuff and wait a while before
taking another reading.
• Correct readings are impossible when diastolic blood pressure is
greater than 180mmHg.
Re-inflation
If the Base Unit determines that pressure in insufficient during a
measurement operation, it automatically re-inflates the cuff. Re-inflation
may also be performed if you move or talk, or if there is excessive noise
or another type of disturbance during a measurement procedure.
• Note that positioning the cuff so its air tube underneath your arm
(especially when your arm is pressing down on the tube) can cause
noise. This can lead to re-inflate or even make proper measurement
impossible.
Stopping a Measurement
To stop an ongoing measurement operation, press the START button.
This causes the Base Unit to deflate the cuff and then go into standby.
Pressing the ON/OFF button also causes the Base Unit to deflate the
cuff and turn off.
Recording Blood Pressure Readings
You can make copies of the "BLOOD PRESSURE RECORD" at the
back of this manual and use them to keep track of changes in your blood
pressure and pulse measurements.
Measurement starting
indicator flashes on the display to
indicator stops
Diastolic Blood Pressure
L
Pulse Rate

ERROR MESSAGES

When any of the errors described below occurs, the Base Unit deflates the
cuff and then waits for further button operation or data communication with
the Watch.
Message
Cause
• The Base Unit cannot
Err 01
detect a suitable pulse.
• PRESET PRESSURE
Err 02
setting is too low.
• Noise due to body
movement.
• The cuff is not connected
Err 03
to the Base Unit properly,
or a hole in the cuff.
• The cuff is too loose on the
arm.
• PRESET PRESSURE
Err 04
value is too low.
• Noise due to body
movement.
• The cuff is too loose on the
arm.
• Increase in cuff pressure
Err 330
due to body movement.
The appearance of the low battery indicator shown below causes the
Base Unit to cancel any ongoing measurement or communication
operation and wait for further button operation.
Indicator
Cause
• Battery power is low.

TROUBLESHOOTING

Check the points noted below whenever you experience operational
problems with the Base Unit.
Problem
• Is battery power low?
Nothing happens when
→ Replace the batteries with a full set of
you press the ON/OFF
button.
new ones.
• Are batteries loaded correctly?
→ Check to make sure that batteries are
loaded correctly.
• Is the cuff's plug connected correctly to the
Inflation is performed, but
the cuff does not inflate.
Base Unit?
→ Check the plug for proper connection.
• Is there a hole in the cuff?
→ Contact the Store/Dealer where
purchased about replacing the cuff.
If the actions recommended above do not correct your problem, the
Base Unit itself may be malfunctioning. In this case, contact the Store/
Dealer where purchased.
BP-1M-5
Action
• Do not move or speak
while measurement is in
progress.
• Make sure the air tube is
not under your arm.
• Check the air tube connec-
tion to the Base Unit.
• Tighten the cuff so there is
no space between your
arm and the cuff.
• If there is air leaking from
the cuff because it is old or
it has a hole in it, replace
the cuff with a new one.
• Do not move or speak
while measurement is in
progress.
• Make sure the air tube is
not under your arm.
• Tighten the cuff so there is
no space between your
arm and the cuff.
• Take care so no pressure
is applied to the cuff while
measurement is in
progress.
Action
Replace the Base Unit
batteries.
What to Check and What to Do