STEP 2
Disconnect the positive, negative wires (RED= VE.) (BLUE=-VE.) then disconnect the bridge
wire (blue) from the + Ve terminal of one battery, and the negative terminal of the other bat-
tery. Batteries can be removed.
STEP 3
To disconnect the circuit breaker, with No 1 Philips screw driver. Remove the cable fixing
screws inside the battery box lid. Use a 9/16" ring spanner to remove the lock nut which
holds the circuit breaker in place.
STEP 4
To remove 4 pin beau socket, use a heavy duty soldering iron to de-solder the wires.
NB make a note or diagram of the colour codes and the connections. Use a Philips screw dri-
ver no. 2 together with an 8mm. Open ended spanner to hold nylock nut at the back of the
beau socket.
STEP 5
To remove the auxiliary power socket. Use a No1 flat blade screw driver to remove the cover
screws. Un-screw the wire terminal fixing screws and pull the wires free of the socket.
NB. Do not disconnect wires from the socket when both batteries are connected, as the
auxiliary feed will still be live.