7. Error Messages
Display/Problem
Display meaning
Measured temperature too high
CAUTION:
High fever requires immediate medical
attention!
Measured temperature too low
Room temperature too high
Room temperature too low
Error function display
Blank display
Dead battery indication
Possible cause and remedy
Displays «H» when measured temperature higher
than 42.2 °C.
Displays «L» when measured temperature lower
than 34 °C.
Displays «H» in conjunction with the «s» when
room temperature is higher than 40.0°C. Lower the
room temperature to below 40°C.
Displays «L» in conjunction with the «t» when
room temperature is lower than 16.0°C. Increase the
room temperature to above 16.0°C or alternatively
hold the thermometer tip in your hands for up to 30
seconds to warm.
When system has a malfunction. Turn off and restart
the thermometer. If problem continues, please
contact our customer service.
Please check if the battery has been loaded
correctly. Also check polarity (<+> and <–>)
of batteries.
If the steady battery icon is the only symbol shown
on the display, the battery should be replaced
immediately.
8
This product conforms to the provisions of the EC directive 93/42/EEC (Medical
Device Directive). This device conforms to the following Standards:
• EN 60601-1 Medical electrical equipment - Part 1: General requirements for safety
• EN 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for safety -
Collateral standard: Electromagnetic compatibility - Requirements and tests
• ASTM 1965-98 Standard Specification for Infrared Thermometers for Intermittent
Determination of Patient Temperature
BF Type applied part.
Consult Instructions for Use
The stipulations of EU-Directive 93/42/EEC for Medical Devices Class IIa have been fulfilled.
0044
8. MemoryTracker
mode
Use the following steps to recall the last 12 temperature measurements.
Display
Pressing and releasing the Start
button after the last 12 readings have been recalled will resume the above sequence from reading 1. The
oldest reading will automatically be replaced by the latest reading. It is not necessary to delete old readings.
Display meaning
Description
Press the button to enter Recall Mode when
Recall Mode
power is off. The memory icon «M» flashes.
Reading 1
Press and release the button to recall the last
(the last reading)
reading. The number 1 will appear showing the
last saved reading.
Reading 2
Press and release the button to recall the second
(the second latest reading)
latest reading.
Reading 12
Press and release the button consecutively to
(the last 12 readings)
recall readings in succession, up to the last 12
readings.