5. Control displays and symbols
LCD display
Display meaning
Self check ensures
proper functioning
Recall of last reading
Ready to measure
Measurement complete
Low battery indication
Description
Press the Start button to turn on the
unit, all segments will be shown for
2 seconds.
The last reading will be shown on the
display automatically for 3 seconds.
The unit is ready for the measurement,
the °C icon will keep flashing.
The reading will be shown on the
LCD display with the °C icon flashing,
the unit is ready again for the next
measurement.
When the unit is turned on, the battery
icon will keep flashing to remind the
user to replace the batteries.
6
6. Directions for use
1. Press the
button on front of unit. The LCD display is activated to show all segments for 2 seconds.
2. The last measurement reading will be shown on the display automatically for 3 seconds with the «M» icon.
3. When the °C icon is flashing, a beep sound is heard and the thermometer is ready for the measurement.
4. Place the ComfortTouch tip gently onto the skin approximately 1.25cm above the center of the eyebrow. If the
eyebrow area is covered with hair, sweat or dirt, please clean the area beforehand to ensure accurate readings.
5. Press the
button once and release it. Start to move the thermometer steadily from 1cm above the center of the
eyebrow to 1.25cm above the tail of the eyebrow area and back up again. A long beep will sound after 3 seconds to
signal the end of the measurement. If the movement is too slow to reach the tail of the eyebrow before the long beep
sound is heard, repeat the measurement as above from step 4, and move the thermometer a little quicker this time.
6. Read the measured temperature from the LCD display.
To avoid an inaccurate measurement, note the following:
• Patients should stay in steady - state room condition for at least 30 minutes.
• Don't remove the measurement device from the measuring area before hearing the final beep.
• Don't take a measurement while or immediately after nursing a baby.
• Don't use the Forehead Thermometer in high humidity environments.
• Patients should not drink, eat, or be physically active before/while taking the measurement.
• Before placing the thermometer sensor onto the measurement area, remove hair, sweat, or dirt.
• Use an alcohol swab to carefully clean the sensor and wait for 5 minutes before taking a measurement on another patient.
• 10 short beeps will sound when the temperature is higher than 37.5 °C in order to alert the patient that he/she may
have a fever.
• Always take the temperature in the same location, since the temperature readings may vary from different locations.
• I n the following situations it is recommended that three temperatures in the same location be taken and the highest one taken as
the reading:
1) Newborn infants in the first 100 days.
2) Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
3) W hen the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the
instrument and obtains consistent readings.