Focus Specifications - AirSep Focus Patient Manual

Focus portable oxygen concentrator
Hide thumbs Also See for Focus:
Table of Contents

Advertisement

Focus Specifications

Oxygen Concentration:*
Pulse Dose:
Dimensions:
Weight:
Power:
Battery duration
(Rechargeable
lithium Battery)
Battery recharge time:
Warm-up time:
Battery cycle life:
Audible alarms and pulse
visual indicators:
**Temperature range:
**Operational Altitude:
* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
**Operating outside of these operational specifications can limit the concentrator's ability to meet Oxygen Concentration specification at higher liter flow rates.
MN170-1 rev C 07/13
Pulse setting equivalent to a continuous flow of 90% oxygen - 3% / +5.5%
17.25ml ± 10%
6.4 in. high x 4.8 in. wide x 2.5 in. deep (16.4 cm high x 12.2 cm wide x 6.1 deep)
Concentrator 1.75 lb (0.8 kg)
Battery 0.53 lb (0.2 kg)
Optional AirBelt Battery 1.8 lb (0.8 kg)
Universal Power Supply:
AC Power Supply: Input # 1_100 – 240 VAC (1.5 Amps max at 120 VAC 50/60 Hz)
DC Power Supply: Input # 2_11-16 VDC 5.0 Amps max
Battery: 1 ½ hours (per battery)
Optional AirBelt Battery: 4 hours
4 hours; optional AirBelt: 3 hours
2 minutes
Approximately 300 cycles, then 80% capacity or below.
Start-up
– audible and visual
Pulse flow
– visual
Cannula disconnect – audible and visual
Breath rate alarm
– audible and visual
General malfunction – audible and visual
Service required
– visual
Low battery
– audible and visual
Battery condition
– battery level indicator on battery
Normal operating temperature: 41°F to 104°F (5°C to 40°C)
Up to 95% RH (non-condensing)
Storage temperature:
-4°F to 140°F (-20°C to 60°C)
Operational Altitude: up to 10,000 ft. (3,048m) (523 mmHg) Higher altitudes may
affect performance
30

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents