FreeStyle Specifications

Oxygen Concentration:*
Dimensions:
Weight:
Power:
Battery duration
(Rechargeable lithium
Battery)
Battery recharge time:
Warm-up time:
Battery cycle life:
Audible alarms and pulse
visual indicators:
Sound:
Temperature range:
Max Operational Altitude:
* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
* Operating unit outside of operational temperature range can affect performance.
MN123-1 rev B 01/10
1-3 pulse settings; equivalent to a continuous flow of 90% oxygen ± 3%
8.6 in. high x 6.1 in. wide x 3.6 in. deep (21.8 cm high x 15.5 cm wide x 9.1 cm deep)
4.4 lb (2.0 kg); 1.8 lb (.8 kg) optional AirBelt
AC Power Supply: 100 – 240 VAC (1.0 amp at 120 VAC)
DC Power Supply: 11-16 VDC (6.0 amps at 12 VDC)
3 LPM – 2 hours; 2 LPM – 2.5 hours; 1 LPM – 3.5 hours
Optional AirBelt when combined with the internal battery:
3 LPM – 5 hours; 2 LPM – 6 hours; 1 LPM – 10 hours
3.5 hours; optional AirBelt: 3 hours
2 minutes
Approximately 300 cycles, then 80% capacity or below.
Start-up – audible and visual (GREEN indicator light); Pulse flow – visual (GREEN
light); Battery condition – battery level (GREEN indicator lights); cannula disconnect-
audible and visual (RED alarm light); System overdraw – audible and visual (RED
warning and alarm lights); High and low pressure – audible and visual (RED alarm
light); Service required: visual (YELLOW alarm light)
38 dBA at 1 setting: 41 dBA at 2 setting: 44 dBA at 3 setting
Operational temperature: 41°F to 104°F (5°C to 40°C)
Storage temperature:
-2°F to 140°F (-20°C to 60°C)
up to 12,000 ft (3,657.6 m) Higher altitudes may affect performance