Filter

Air Intake Filter
At least one time each week, remove the cover on the lower front of the unit, and wash the air intake filter, which is
positioned on the inside of the cover. Your Equipment Provider may advise you to clean it more often, depending upon
your operating conditions. Follow these steps to properly clean the air intake filter:
1. Remove the filter from the air intake filter cover, and wash it in a warm solution of soap and water.
2. Rinse the filter thoroughly, and remove excess water with a soft, adsorbent towel. Ensure that the filter is dry
before replacing it.
3. Replace the dry filter.
Do not operate FreeStyle without the air intake filter in place. If a second filter is provided, insert the "replacement" filter
before you clean the dirty filter.
AirSep does not recommend the sterilization of this equipment.
MN123-1 rev B 01/10
22