14. Troubleshooting

Symptom
Possible Cause
Bath lift does not
• Battery is flat.
move. No sound
• B attery is not
from the motor.
connected to hand
control.
• N ot pressing correct
area on hand
control buttons.
• F aulty electronics in
hand control.
• Battery fault.
• Actuator fault.
Red light on hand
• B attery is flat or
control is on and
has not been fully
bath lift only
charged and Stop
moves up.
Descent™ has been
activated to prevent
the bath lift lowering.
Red light on hand
• B attery is flat and has
control is on but
not been charged for
bath lift does not
long period (typically
move.
7-10 days).
• Battery fault.
Solution
• Charge battery.
• I nsert battery correctly into both tracks
on hand control cradle and slide all the
way to the end until it clicks.
• P ress arrowhead on either UP or DOWN
buttons. Continue to press.
• C ontact supplier or authorized service/
repair center.
• C ontact supplier or authorized service/
repair center. Replace battery.
• C ontact supplier or authorized service/
repair center.
• Charge battery immediately.
• Charge battery immediately.
• C harge battery before use if bath lift has
not been used for long periods (typically
7-10 days).
• D o NOT start bath with RED light on
hand control. Charge immediately.
• E nsure battery is placed on charge as
soon as you finish your bath if FLASHING
GREEN light appears on hand control.
• W hen bath lift is not in use, leave battery
connected to hand control. Disconnect
only for charging or when moving bath lift.
• I t is recommended the battery be charged
after each bath, wherever possible.
• C ontact supplier or authorized service/
repair center.
Section
Section 2
Section 5
Section 5
n/a
n/a
n/a
Section 2
Section 2/
Section
5 &
Section 7
n/a
15

Hide quick links:

Advertisement

Table of Contents
loading