The monitoring system will not work properly at altitudes greater than 8,000 feet.
The System is designed to use at temperatures between 57.2°F and 104°F (14°C and 40°C)
and below 85% of relative humidity. If you use the system outside the proposed conditions, it
can give false results.
Be sure to store the test strips between 40°F and 86°F (4°C and 30°C) and avoid direct
sunlight.
Be sure to follow your local regulations for proper disposal of used test strips and lancets.
Do not use this meter in a dry environment, especially if synthetic materials are present.
Synthetic clothes, carpets, etc., may cause damaging static discharges in a dry environment.
Do not use this meter near cellular or cordless telephones, walkie talkies, garage door
openers, radio transmitters, or other electrical or electronical equipment that are sources of
electromagnetic radiation, as these may interfere with the proper operation of the meter.