AirSep FreeStyle Service Manual page 62

Portable oxygen concentrator
Hide thumbs Also See for FreeStyle:
Table of Contents

Advertisement

Problem
Battery duration is less
than per specification.
Note: While charging, battery
may not initiate a charge.
Allow unit to cool to operating
temperature before charging.
Will initiate in approx. (20
min).
Note: Whenever storing
FreeStyle for an extended
period of time, ensure that the
battery the state of charge
(SOC) is not depleted or
stored at 100%. Both these
conditions can result in
unrecoverable capacity loss.
The recommend SOC is 25%
to 50% for long-term storage.
The recommended storage
temperature is 23º C ± 2º C.
Unit does not start
on battery.
Battery does not charge.
AirSep Corp. — FreeStyle Service Manual
Possible Cause
Battery not fully charged.
Battery has been stored in a
discharged or 100% charged
state for prolonged period of
time or stored in a very high
temperature environment.
Battery reached cycle life.
Unit temperature has exceeded
battery operating temperature
5 to 40ºC.
Battery reached cycle life
Battery is defective
Unit temperature has exceeded
battery operating temperature
5 to 40ºC.
Battery is defective.
AC adapter connector is
defective or contact pin is bent.
Defective AC adapter
Circuit Board
Solution
Ensure that battery
charges for at least
3½ hours and all (4) LEDs
are lit solid.
Replace battery.
Replace battery.
Allow unit to cool
before starting.
Replace Battery
Allow battery to cool before
charging.
Replace battery
Replace battery.
Replace AC adapter.
Replace main circuit
board
7-3