Advertisement

5.15 Control Panel/Key pad Replacement
When replacing the control panel keypad, always use and ESD safe
work area. Also, use an ESD wrist strap to protect electronic circuitry.
1. Remove the front and rear cover per Section 5.2.
2. Remove PSA system per Section 5.4.
3. Remove control panel/keypad flex circuit from J8 Ziff (zero insertion force)
4. Remove the flex tab circuit from rib guides, and peel off the control panel /
5. Replace control panel/keypad with new control panel keypad.
6. Check control recessed area of enclosure center section to ensure plastic is
7. Remove adhesive backing and seal backing from new control
AirSep Corp. — FreeStyle Service Manual
connector by carefully lifting upward on locking tab.
keypad from the center enclosure. (Figure 5.40)
clean and all adhesive backing and seal backing from new control panel
keypad.
panel/keypad. (Figure 5.41)
Figure 5.40
Figure 5.41
5-27