Operating and Installation Instructions
1

Advertisement

Table of Contents
loading

Summary of Contents for OHAUS CHAMP II BASE

  • Page 1 CHAMP II BASE Operating and Installation Instructions...
  • Page 2 Ohaus Corporation, 19A Chapin Road, P.O. Box 2033, Pine Brook, NJ, 07058, USA Declaration of Conformity We, Ohaus Corporation, declare under our sole responsibility that the instrument models listed below marked with “CE” - are in conformity with the directives and standards mentioned.
  • Page 3 NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITH THE LIMITS. FOR A CLASS A DIGITAL DEVICE, PURSUANT TO PART 15 OF THE FCC RULES. THESE LIMITS ARE DESIGNED TO PROVIDE REASONABLE PROTECTION AGAINST HARMFUL INTERFERENCE WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQUIPMENT GENERATES, USES, AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NOT INSTALLED AND USED IN ACCORDANCE WITH THE INSTRUCTION MANUAL, MAY CAUSE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS.
  • Page 4: Champ Ii Base

    CHAMP II BASE 1. Overview CHAMP II is a single load cell base. This scale base can be connected to CD-11, CD-31 or CD-33 Ohaus Industrial Indicator (or other types). 2. Physical Descriptions Construction Platform - Fabricated stainless steel Scale Frame - Formed and welded mild steel, painted black.
  • Page 5: Installation Procedures

    3. Installation Procedures Examine the shipping box for any signs of damage. If DAMAGE is found, make a claim with the carrier immediately. Open the box and remove the unit. Place the unit on a stable flat surface. Connect to the indicator. Level the scale, apply power and weigh.
  • Page 6: Troubleshooting

    4. Troubleshooting If operational difficulties are encountered, first obtain as much information as possible regarding the problem. Failures and malfunctions often may be traced to simple causes such as loose connections to the indicator, low battery power, improper setup, etc. If simple causes cannot be found, additional troubleshooting is best performed by substitution.
  • Page 7 Overload Stop Adjustments The overload stop gaps must be checked and reset if the top or bottom frame, or load cell is replaced. To set the gaps, remove the platform, loosen the jam nuts (refer to Figure 1-1), then use the proper size feeder gauge in the gap, turn the set screws until a slight drag on the feeder is felt.
  • Page 8 Shift Test A shift test verifies that all sections of the scale platform weigh within tolerance. If the scale does not pass the shift test, verify the overload stop gaps before replacing the load cell. No adjustment for the shift is possible. If the shift test cannot be passed, the load cell must be replaced.
  • Page 9: Parts List

    Parts list OHAUS PART NO. DESCRIPTION 71129642 Indicator Bracket Assembly 71125105 Stainless Steel Column, 350 mm high 71125103 Stainless Steel Column, 680 mm high 71129064 Column Bracket 71133398 Load Cell, MT1241-30kg, CH15R 71123975 Load Cell, MT1241-50kg, CH30R 71123976 Load Cell, MT1241-100kg, CH60R, CH60L...
  • Page 10: Limited Warranty

    Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.