Carry A Daily Kit; Testing Your Blood Glucose; Perform A Control Solution Test - Abbott FreeStyle Aviator User Manual

Insulin delivery and glucose monitoring system
Table of Contents

Advertisement

Section 5: Testing Your Blood Glucose

5.1 Perform a Control Solution Test

4.3 Carry a Daily Kit

Purpose
To take care of your daily pump needs.
When
Do this procedure every day.
Materials
Keep a daily kit with you at ALL times.
This kit should include:
Blood glucose monitoring supplies
Ketone monitoring supplies
Glucose tablets/gel or liquid or other source of carbohydrate to treat a low glucose condi-
tion.
A Glucagon Emergency Kit (Glucagon Emergency Kits have an expiration date).
Important Reminder: Check frequently to make sure your Glucagon Emergency Kit has NOT
expired.
Extra infusion sets
Insulin
Extra unopened pump reservoirs
An extra AAA alkaline battery
Dressings and adhesives
Spare battery cap
Insulin syringe and short or rapid-acting insulin vial or insulin pen.
Important Reminder: Insulin is sensitive to temperature extremes. Replace your spare insulin
monthly.
DOC15736_Rev-A-Man.indd 73-74
73

5 Testing Your Blood Glucose

5.1 Perform a Control Solution Test
Purpose
To check that the integrated FreeStyle Lite Blood Glucose Meter and FreeStyle Lite Test
Strips are working properly without using any blood.
When
Do this procedure:
Aviator Companion, before doing a blood glucose test with the
When you first get your
integrated FreeStyle Lite Blood Glucose Meter.
Whenever you need to practice testing without having to use your own blood.
You want to confirm that your companion and test strips are working properly.
Materials
FreeStyle Lite Test Strips
FreeStyle Control Solution
Your
Aviator Companion fully operating
Helpful Information
The FreeStyle Control Solution is a red liquid that contains a fixed amount of glucose.
For information on how to obtain control solution, call Customer Care.
Important: FreeStyle Control Solution is good for three months after opening the bottle or until the expiration date
printed on the label, whichever comes first.
Important: Count forward three months from the date you open a new bottle of control solution. This is your discard
date. Write this date on the side of the control solution bottle.
Example: You open the control solution on January 15
date that you record on the bottle.
Section 5: Testing Your Blood Glucose
5.1 Perform a Control Solution Test
. Count forward three months to April 15
. This is the discard
th
th
74
2/9/09 11:29:39 AM

Advertisement

Table of Contents
loading

This manual is also suitable for:

FreestyleAviator

Table of Contents