Problem in Operation

1. If the meter does not display a message after inserting
a test strip:
PROBABLE CAUSE
a.Battery exhausted.
b.Battery incorrectly
installed or absent. .
c.Test strip inserted upside
down or incompletely.
d.Defective meter.
2. If the test does not start after applying the sample:
PROBABLE CAUSE
a.Insufficient blood sample.
b.Defective test strip.
c.Sample applied after
automatic shutoff (three
minutes after last user
action).
d.Defective meter.
44
WHAT TO DO
Replace the Battery.
Check that the battery is correctly
installed with the positive "+" side up.
Insert the test strip correctly with the
contact bars end first and facing up.
Please call your local customer
service for help.
WHAT TO DO
Repeat the test with a new test strip
and a larger sample.
Repeat the test with a new test strip.
Repeat the test with a new test strip;
apply sample only when "
on the display.
Please call your local customer
service for help.
" appears