Quickie Rhythm Technical & Service Manual page 68

Table of Contents

Advertisement

QUICKIE ELECTRONICS SECTION
Actuator Test (QMAC controlled actuators)
Complete the following steps to test the seating system
actuators.
1.
Connect the Quickie iQ or HHP to the wheel-
chair.
2.
Turn on the wheelchair. Prior to completing any
test, re-home the wheelchair using either the HHP
or the PCSS.
3.
If using the HHP, select Monitor, then QMAC Ac-
tuators. Under QMAC Actuators, select the actua-
tor to be tested (i.e., Tilt). (fi g2.23.26)
4.
If using the Quickie iQ, select Monitor Parameter,
followed by selecting QMAC, then actuator to test
(i.e., Tilt). (fi g 2.23.27)
5.
Note the position of the actuator (counts and
degrees). Operate the actuator and observe the
counts, degrees and current. Refer to Table 1 for
a listing of typical currents for each actuator.
Fig 2.23.27
Actuator
Recline
Tilt
Lift
Left Leg
Right Leg
Power Shear
Table 1
©2006 Sunrise Medical
Fig 2.23.26
Average Current – No Load
1 to 2 amps
1 to 2 amps
1 to 2 amps
1 to 2 amps
1 to 2 amps
1 to 2 amps
68
Tilt
+ + -
Angle
Counts
Current
0.000
SW2 Status
Maximum Current –Under Load
8 amps
8 amps
8 amps
4 amps
4 amps
4 amps
3.0
deg
0
A
open

Hide quick links: