QUICKIE ELECTRONICS SECTION
Actuator Test (QMAC controlled actuators)
Complete the following steps to test the seating system
actuators.
1.
Connect the Quickie iQ or HHP to the wheel-
chair.
2.
Turn on the wheelchair. Prior to completing any
test, re-home the wheelchair using either the HHP
or the PCSS.
3.
If using the HHP, select Monitor, then QMAC Ac-
tuators. Under QMAC Actuators, select the actua-
tor to be tested (i.e., Tilt). (fi g2.23.26)
4.
If using the Quickie iQ, select Monitor Parameter,
followed by selecting QMAC, then actuator to test
(i.e., Tilt). (fi g 2.23.27)
5.
Note the position of the actuator (counts and
degrees). Operate the actuator and observe the
counts, degrees and current. Refer to Table 1 for
a listing of typical currents for each actuator.
Fig 2.23.27
Actuator
Recline
Tilt
Lift
Left Leg
Right Leg
Power Shear
Table 1
©2006 Sunrise Medical
Fig 2.23.26
Average Current – No Load
1 to 2 amps
1 to 2 amps
1 to 2 amps
1 to 2 amps
1 to 2 amps
1 to 2 amps
68
Tilt
+ + -
Angle
Counts
Current
0.000
SW2 Status
Maximum Current –Under Load
8 amps
8 amps
8 amps
4 amps
4 amps
4 amps
3.0
deg
0
A
open