Internal Calibration

If Balance is connected externally to PC or Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated after
the completion of Calibration process.
Successful Calibration
When GLP ON
------------------------
20-Jul-10
Model
Ser.no.
Ver.no.
ID
------------------------
Calibration:
Start:
Temperature
Diff.
Internal Cal Done
Diff.
------------------------
20-Jul-10
Name:
------------------------
When GLP OFF
Calibration:
Start:
Temperature
Diff.
Internal Cal Done
Diff.
4.2 Internal Calibration
To obtain weighing results, the balance must be matched to the acceleration due to
gravity at its location.
Calibration is necessary
ð
Before the balance is used for the first time.
ð
At regular intervals during weighing operation.
ð
After a change in location.
Procedure
To obtain accurate results, the balance must be connected to the power supply and
allowed to warm up to the operating temperature as described on Page No
Ensure that the weighing pan is unloaded and close the doors of the draft shield (if
used). Balance should be Zero before calibration.
ð
Press and hold <CAL> key, display, shows
"CAL Int"
ð
Release <CAL> key now.
Internal Calibration process Starts.....
- When the Internal Weight is being loaded "C" will be displayed on display.
- When the Internal Weight is being unloaded "CC" will be displayed on display.
Calibration is finished when 'Int.done' is message is displayed.
The adjustment can be terminated at any time using the <CANCEL> key.
The following message appears : 'Abort'
Internal Calibration Report
10:32AM
Citizen
CX
220
9223102
r0.1.5.3
1234567
Internal
Manual
29.449'C
+
0.0009g
0.0000g
10:32AM
................
Internal
Manual
29.449'C
+
0.0009g
0.0000g
25
Unsuccessful Calibration
When GLP ON
--------------------------------
20-Jul-10
10:34AM
Citizen
Model
CX
Ser.no.
9223102
Ver.no.
r0.1.5.3
ID
1234567
------------------------
Calibration:
Internal
Start:
Manual
Temperature
29.495'C
Internal Cal Failed
------------------------
20-Jul-10
10:34AM
Name:
................
------------------------
When GLP OFF
Calibration:
Internal
Start:
Manual
Temperature
29.495'C
Internal Cal Failed
26