In critical applications where very
stringent requirements have to be met,
it is strongly recommended that all the
relevant factors for the process, such
as detergent, quality of water, etc. are
discussed with a Miele Application
Technology specialist.
~
It is the responsibility of the operator
to routinely check that the required
cleaning standards are being met.
Therefore, results need to be regularly
tested and documented, both thermo
electrically and through inspection. For
thermo chemical processes, additional
testing is required using chemical or
biological indicators.
~
The performance of the cycles was
tested using Neodisher products. The
use of other products is not
discouraged but may not yield identical
results. Please be aware that changes
in formulation, storage conditions, etc.
which may not be publicized by the
chemical manufacturer can have a
negative effect on cleaning results. and
are not the responsibility of the machine
manufacturer.
IMPORTANT SAFETY INSTRUCTIONS
~
specialized products, it is essential that
the manufacturer's instructions are
followed. Only use the product for the
application described by the
manufacturer to avoid any material
damage or the occurrence of strong
chemical reactions.
~
special inserts should only be used for
their specific applications. Instruments
with hollow parts must be flushed
through with wash water.
~
before arranging them in the machine.
~
solvents or corroding ferrous material,
and in particular hydrochloric acid or
chloride solutions to get into the wash
cabinet. The presence in compounds of
any solvents should be minimal,
(especially those in hazard class A1).
~
ensure that solutions or steam
containing hydrochloric acid do not
come in contact with the stainless steel
casing of the machine.
~
in these instructions and the separate
Installation Instructions.
Please note the following symbols on
the machine:
When using cleaning agents and
The mobile units, baskets and
Empty all containers and utensils
Do not allow any remains of acids,
To avoid any corrosion damage
Please follow the installation advice
Caution : See Operating
Instructions!
Caution : Danger of electric
shock!