Following termination of ventilation the
Oxylog is to be prepared for thorough
cleaning and disinfection:
Remove mask or tube from patient
connection of non-rebreathing valve.
Detach ventilation tubing from non-
rebreathing valve and from Oxylog.
Disassemble non rebreathing valve into
its component parts (Fig. 10). Turn
cover 45" in anti-clockwise direction
with respect to valve housing (= valve
open).
Caution!
The red check valve must not be
removed from the yellow control valve!
The PEEP valve is to be disassembled
into its 3 main components.
The disassembled parts of the patient
system are to be thoroughly cleaned
either in running hot water or in a
mixture of detergent' )/water with sub-
sequent rinsing under running water.
Important: Do not use a hard brush for
cleaning purposes!
Perfect (Henkel Co ) and Caraform (Welgert Co.)
The surface of the device can be
cleaned using a soft cloth impregnated
with a mixture of detergent/water.
Caution!
Do not use petrol, ether or similar
solvents for cleaning the device.
Carefully rinse Volumeter 3000 in
direction of flow with hot running water.
Caution!
Do not allow water to flow into drain
holes in control section.
Carefully remove residual water from
Volumeter.
All parts are to be thoroughly dried
following cleaning and rinsing.
The component parts can also be
washed in the Drager Purfactor@, which
automatically subjects the material to
be washed to disinfection and drying.
Disinfection in liquid disinfectant
The cleaned and dried parts of the
patient system can be disinfected in a
cold disinfectant solution.
The surface of the device can be
disinfected by wiping over it.
Caution!
Use may only be made of disinfectants
which do not attack the materials used.
Compliance with the prescribed con-
centrations is to be ensured. In case of
doubt consult the disinfectant manu-
facturer.
Following disinfection and drying of its
component parts, the Oxylog is to be
assembled as described under aylnitial
to a functional check as described
under a>Functional Check<< on Page il.
Disinfection in Drlger Aseptor@
The cleaned and dried components of
the patient system as well as the device
itself can be disinfected in the Aseptor.
Caution!
The ventilation tubing and breathing
masks (silicone rubber) must not be
disinfected in the Aseptor.
Following disinfection in the Aseptor,
t h e d e v i c e i s t o b e a s s e m b l e d a s
described under a,lnitial Preparation<<
on Page 8 and subjected to a functional
check as described under b,Functional