DIGITAL
MULTIMETER
INSTRUCTION MANUAL
KU-2600
FOR SAFETY MEASUREMENTS!!
To prevent an electrical shock hazard to the operator and/or damage to the
instruments, read this instruction manual carefully before using the Digital
Multimeter. WARNINGS with the symbol
on the Digital Multimeter and
this instruction manual are highly important.
Important Symbols
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution
(refer to instruction manual)".
WARNING : The symbol in this manual advises the user of an
electrical shock hazard that could result in serious injury or even
death.
CAUTION : The symbol in this manual advises the user of an
electrical shock hazard that could cause injury or material damages.
WARNING
KU-2600 is designed to comply with CAT ‡U 600V. But, do not measure High
Power Line of more than 3kVA power. High Power Line sometimes includes
High Surge Voltage that could possibly induce dangerous arcs of explosive
short in the instrument and could result in serious injury to the operator. Even
if it is Low Power Line, use extreme care when measuring high voltage.
INTRODUCTION
Thank you for purchasing KAISE "MODEL KU-2600 DIGITAL MULTIMETER".
To obtain the maximum performance of this instrument, read this Instruction
Manual carefully, and take safe measurement.
1. UNPACKING AND INSPECTIONS
Inspect the instrument and acessories for transport damage. If any damage or
missing items are found, ask your local dealer for replacement.
Confirm that the following items are contained in the package.
1. Digital Multimeter and Test Leads
2. Two 1.5V R6P Batteries (Installed and spare)
3. Two Spare Fuses (0.5A/250V, 10A/250V)
4. Holster
5. Instruction Manual
2. SPECIFICATIONS
2-1. GENERAL SPECIFICATIONS
1. DISPLAY (LCD)
a. Numerical Display : 4000 count LCD, 12mm high
b. Units and Symbols :
mV, V, Hz, •", ƒ¶, kƒ¶, Mƒ¶, nF, ƒÊF,
, •@, mA, A,
REL•¢, •@ , H , OL, AUTO, AC, DC, •| and decimal point.
•| •{
2. OPERATING PRINCIPLE :
ƒ°
‡™ conversion
3. RANGE SELECTION : Auto-Ranging / Range Hold
4. SAMPLING RATE : 2 times per second
5. POLARITY : Auto-Polarity ("•["symbol appears in minus)
6. OVERRANGE INDICATION : "OL" symbol appears. (excluding DC/AC 600V)
7. DIFFERENCE MEASUREMENT (Zero Adjustment) : Press REL•¢ Key to
make Difference Measurement and Zero Adjustment.
8. DISPLAY HOLD : Press HOLD Key.
9. BATTERY WARNING :
symbol appears when battery voltage goes down
•| •{
below approx. 2.4V.
10. OPERATING TEMPERATURE & HUMIDITY : 0•Ž to 40•Ž, less than 75•" RH
in non-condensing.
11. STORAGE TEMPERATURE & HUMIDITY : •|20•Ž to 60•Ž less than 80•"
RH in non-condensing.
12. POWER SUPPLY : Two 1.5V R6p Batteries.
13. POWER CONSUMPTION : 4.5mW typically.
1
14. BATTERY LIFE : 150 hours continuous operation.
15. AUTO POWER OFF : Power turns off automatically in 15 minutes after any
switch operation.
16. DIELECTRIC STRENGTH : 3.7kV rms for one minuite between input
Terminal and Cases.
17. FUSES :
mA Function : Fast Active 0.5A/250V, 5•~20mm
10A Function : Fast Active 10A/250V, 6•~30mm
18. OVERLOAD PROTECTION :
a. V : 1000V DC or 750V AC rms max. for 1 minute.
(400mV Range is 500V rms )
b. ƒ¶/•@ /
/
: 250V rms max. for 1 minute.
19. DIMENSIONS & WEIGHT : 155•~75•~29 mm, 180g
20. SAFETY LEVEL : IEC-61010-1, Overvoltage CAT.‡U 600V, CAT‡V300V.
21. ACCESSORY : One Pair of Test Leads, Batteries (Installed), Carrying Case,
Instruction Manual
22. OPTIONAL ACCESSORY : 940 Alligator Clips.
2-2. MEASUREMENT SPECIFICATIONS
(23•Ž•}5•Ž•A•ƒ80•"RH in non-condensing)
1. DC Voltage ( •@ V )
Input
Max Input
Range
Accuracy
Resolution
Impedance
Voltage
400.0mV
100ƒÊV
•†100Mƒ¶
4.000 V
•}0.5%rdg•}4dgt
1mV
600V DC
40.00 V
10mV
•à10Mƒ¶
400.0 V
100mV
•}1.0%rdg•}4dgt
600 V
1 V
2. AC Voltage ( •`V )
Input
Max Input
Range
Accuracy
Resolution
Impedance
Voltage
400.0 mV
•}4.0%rdg•}8dgt
0.1 mV
4.000 V
1mV
600V
40.00 V
•}1.2%rdg•}5dgt
10mV
•à10Mƒ¶
AC rms
400.0 V
100mV
600 V
•}1.5%rdg•}5dgt
1 V
Frequency Response : 50Hz~500Hz. 500Hz~1000Hz : •}3.5%rdg•}5dgt.
*400.0mV Range : 5mV~400mV.
3. Frequency ( Hz )
Input
Max. Input
Range
Accuracy
Resolution
Sensitivity
Voltage
0.01Hz
9.999Hz
•}0.2%rdg
250V
1V RMS
•`9.999MHz
•}1dgt
•`1.0kHz
AC rms
4. Duty Cycle ( •" )
Input
Max. Input
Accuracy
Resolution
Range
Sensitivity
Voltage
0.0•"
250V
•}0.2%rdg•}2dgt
0.1•"
1V RMS
•`99.9•"
AC rms
Frequency Scope : 1Hz•`1kHz
5. Resistance ( ƒ ¶ )
Open Circuit
Overload
Range
Accuracy
Resolution
Voltage
Protection
400.0 Ħ
0.1 Ħ
1 Ħ
4.000kĦ
•}1.0%rdg•}3dgt
250V
10 Ħ
40.00kĦ
•à0.44V
DC/AC rms
400.0kĦ
100 Ħ
1 kĦ
4.000MĦ
•}1.8%rdg•}3dgt
10 kĦ
40.00MĦ
6. Diode Tests (
)
Test
Open Circuit
Overload
Accuracy
Range
Current
Voltage
Protection
250V
•}5.0%rdg•}3dgt
4.000V
•...0.5mA
DC/AC rms
8. Capacitance ( •@ )
Test
Range
Accuracy
Resolution
Current
•}5.0%rdg•}5dgt
0.1nF•`40nF
0.001nF•`nF
40nF•`40ƒÊF
nF•`ƒÊF
•}3.0%rdg•}5dgt
•à1.2V
•}4.0%rdg•}5dgt
40ƒÊF•`200ƒÊF
ƒÊF
Overload Protection : 250V DC/AC rms
9. DC Current (
mA /
A )
Voltage
Accuracy
Resolution
Max. Input Voltage
Range
Drop
40.00 mA
10ƒÊA
200mV
•}2.0•"rdg•}4dgt
400mA
400.0 mA
100ƒÊA
max.2.0V
10.00 A
•}2.5•"rdg•}4dgt
10 mA
1.0V
10A •iwithin15sec.•j
NOTE :
40.00mA, 400.0mA : 0.5A/250V Fuse Protected
10.00A : 10A/250V Fuse Protected
2
10. AC Current ( •`mA / •`A )
Voltage
Range
Accuracy
Resolution
Max. Input Voltage
Drop
40.00 mA
•}2.5•"rdg•}5dgt
10ƒÊA
200mV
400mA
400.0 mA
100ƒÊA
max.2.0V
•}3.0•"rdg•}5dgt
10.00 A
10A •iwithin15sec.•j
10 mA
1.0V
NOTE :
40.00mA, 400.0mA : 0.5A/250V Fuse Protected
10.00A : 10A/250V Fuse Protected
3. NAME ILLUSTRATION
LCD
KU-2600
REL•@ Key
RS232
AUTO
REL
H
•| •{
RANGE Key
AC
KMĦH
DC
z
•"
nmFVA
HOLD Key
SELECT Key
SELECT
RANGE
REL
HOLD
Hz / DUTY Key
RESET Key
Auto Power OFF
RESET
Hz/DUTY
Hz
V
A
•`
m
•`
V
A
•`
FUNCTION SWITCH
OFF
OFF
10A Terminal
COM Terminal
10A
mA Terminal
COM
VĦHz
VĦHz Terminal
FUSED
10A/15sec MAX
CAT.
mA
(•@•@•@•@ )
300V
•`
FUSED
MAX
500mA MAX
600V •@ MAX
•`
Test Leads
Test Prods
Test Plugs
1. LCD
REL
H
RS232
AUTO
•| •{
AC
KMĦH
DC
z
•"
nmFVA
DC (•@)
•F
Direct Current
AC (•`)
•F
Alternative Current (AC)
•|
•F
Minus Symbol when polarity is minus
•F
Units of Voltage
mV, V
Hz
•F
Unit of Frequency
•"
•F
Unit of Duty Cycle
•F
Units of Resistance
Ħ, kĦ, MĦ
nF, ƒÊF
•F
Units of Capacitance
•F
Diode Tests
•F
Continuity Tests
•F
AUTO
Auto-ranging
H
•F
Display Hold
•F
Difference Measurements
REL
•F
Battery Warning Symbol
•| •{
2. FUNCTION Switch
Set FUNCTION Switch to desired position.
3. SELECT Key (
/
or DC/AC selection)
Select
or
, and DC or AC.
4. RANGE Key
When making voltage, resistance or mA measurements, press RANGE Key to
select Range Hold. AUTO symbol turns off from LCD. There are two ways in
Range Hold.
1.
Just press RANGE Key several times to select a desired range watching the
position of decimal point. The decimal point moves from the lowest range to
the highest range and circulates with each press.
2.
When taking voltage, resistance or mA measurements, press RANGE Key.
The range to which the input value belongs is held and AUTO symbol turns off.
To cancel Range Hold, press RANGE Key for more than 1 second and AUTO
symbol turns on.
To cancel Range Hold, press RANGE Key for more than 1 second and AUTO
symbol turns on.
5. REL•¢ Key
When measuring capacitance, press REL•¢ Key to make zero adjustment. In the
other functions, (•` V,
V, Ħ, mA, A) press this Key to make difference
measurements.
To cancel this Key, press it again and REL•¢ symbol disappears.
6. HOLD Key
Press HOLD Key while measuring, a measuring value is held on LCD with H
symbol shown.
To cancel this Key, press it again and H symbol disappears on LCD.
3
7. RESET Key
Press this Key to reset the operation of the function. This Key is used when
operation is mistaken in the measurment.
8. Hz/Duty Key
Press this Key to select Hz or Duty on •`V,
V, Hz, mA and A positions.
4. SAFETY PRECAUTIONS
Correct knowledge of electric measurements is essential to avoid unexpected
danger such as operator's injury or damage to the instrument. Read carefully and
observe the following precautions for safety measurements.
4-1. WARNINGS
WARNING 1. Checks of Body and Test Lead
Before measurement, confirm the body of this instrument and handle insulators of
the Test Leads have no cracks or any other damages. Dust, grease and moisture
must be removed.
WARNING 2. Warning for High Power Line Measurements
High Power Line (High Energy Circuits) such as Distribution Transformers, Bus
Bars and Large Motors are very dangerous. High Power Line sometimes includes
High Surge Voltage that could cause explosive short in the instrument and could
result in shock hazard. For voltage measurement of High Power Line, do not
touch Clamp Meter, its Test Leads, and any part of the circuit.
WARNING 3. Warning for High Voltage Measurements
Even for Low Energy Circuits of electric/electronic appliances, such as heating
elements, small motors, line cords and plugs, High Voltage Measurements are
very dangerous. Do not touch Clamp Meter, Test Leads, and any part of the
circuit. Generally, shock hazard could occur when the current between the circuit,
fig.1
that involves more than 33V rms or 46.7V DC or peak, and ground goes up to
0.5mA or more.
WARNING 4. Dangerous Voltage Measurement Procedure
For dangerous voltage measurement, strictly observe the following procedure.
1.
Before measurement, turn off the power of the circuit to be measured.
2.
Plug Black Test Lead into COM terminal and Red Test Lead into VĦHz
terminal.
3.
Attach Black and Red Alligator clips (optional) to Test Lead pins.
4.
Set FUNCTION Switch to •`V or
V.
Confirm the power of the circuit to be measured is turned off. Connect Black
5.
Alligator Clip to •| (earth) side and Red Alligator Clip to + (positive) side.
(Refer to fig.2)
Place the Digital Multimeter away from your body not holding it in your hands.
6.
Take safety distance from power source or circuit not to touch the dangerous
voltage.
Turn on the power of the circuit. Read the measurement value on LCD.
7.
8.
After measurement, turn off the power of the circuit and discharge the all
capacitors.
Disconnect Alligator Clips (with Test Leads) from the circuit.
9.
In case of live-line measurement, observe the following procedure.
1.
Place the Digital Multimeter away from your body not holding it in your hands.
2.
Set FUNCTION Switch to •`V or
V.
3.
Attach Black Alligator Clip to Black Test Lead pin. Connect it to •| (earth) side
of the circuit to be measured.
4.
Take safety distance from power source or circuit not to touch the dangerous
voltage.
5.
Hold Red Test Lead with one hand and connect it to •{ (positive) side. (Refer
to fig. 3)
6.
Read the measurement value on LCD.
7.
After measurement, disconnect Red Test Lead and Black Alligator Clip from
the circuit.
WARNING 5. Maximum Input Observance
Do not attempt to measure voltage that might exceed 600V AC or DC, the
specified maximum input of this instrument.
KU-2600
SELECT
RANGE
REL
HOLD
Auto Power OFF
RESET
Hz/DUTY
Alligator
Hz
V
m
A
•`
•`
V
A
•`
Clip
OFF
OFF
10A
COM
FUSED
10A/15sec MAX
Test Prod
CAT.
mA
300V
•`
FUSED
MAX
500mA MAX
600V •@ MAX
•`
fig.2
WARNING 6. Correct Selection of FUNCTION Switch
When taking measurement, always confirm that FUNCTION Switch is set to
correct position. Do not measure voltage on Ħ,
4
Red
Test Prod
Alligator
Clip
fig.3
,
,
, mA and A positions.