Remove the detector from the dock. The
detector is now fit for use.
Once the detector is removed from the
dock, the dock will reset itself and prepare
itself for the next instrument. In
approximately 10 seconds the display will
show the READY screens. The dock is
then ready for the next detector.
4.2

IrDA Connection Issues

4.2.1 Connection Lost

In the event of a loss of communication
between the dock and the instrument, the
LCD will first indicate that the IrDA is not
responding.
If the dock fails to reestablish the
connection, the LCD will indicate
"Connection Lost".
Remove the instrument from the dock.
The dock should reset itself in
approximately 10 seconds.
"IrDA Busy" may also be shown.

4.2.2 Outright Failure to Connect

If the MultiPro is placed in the dock and
connection fails outright, a dot will appear
on the screen to the right of the date. The
dot is an indicator that the instrument IS in
the dock, but the dock is unable to
establish any type of connection to it.
The dot will disappear as soon as the
dock establishes connection with the
dock.
If the dot persists, take the MultiPro out of
the dock and turn it on before returning it
to the dock. If that fails to correct the
problem, call Honeywell Analytics for
more information.

4.3 Test Failures

Failures can occur for a number of
reasons. When a fault occurs, the reason
for the failure will be shown on the dock's
LCD. If the Express Software is running
and the dock is properly connected to the
PC, the fault will also be indicated in the
output box on the instrument tab in the IQ
Express software.

4.3.1 No gas detected

If the detector fails to respond to the
calibration gas, the message "NO GAS
DETECTED" will be shown.

4.3.2 Span calibration failure

If the IQ Express is unable to complete
the span calibration, the dock will show
the failure on the LCD. The failure will
also be shown in the output box on the
instrument page in the PC software.

4.3.3 Alarm test failure

If the dock fails to recognize an alarm that
should be present, it will indicate it on the
display screen following the alarm test.
The failure will also be indicated in the
software in the instrument information.

4.3.4 PC Aborted Test

If the PC is unable to complete the tests
for any reason, the following screen will
be shown:
This message may be shown for a variety
of reasons. See the Calibration Status
output box for further details.