The output box is updated as the
instrument is detected, connected, and
processed. When faults occur, the details
of the fault will be shown and a red fault
light will be shown. Below, a span
calibration failure is the cause of the fault.
If the datalogger has been downloaded,
the "View Session Data" control will
appear on the right. Click on "View
Session Data" to directly access session
information for the instrument using the
BioTrak Program.
If the dock is currently empty, none of the
status indicators will be lit.

3.3 Advanced Settings / Passcode

The MultiPro IQ Express Dock may be set
to verify a specific 4-digit passcode (PIN)
in any instrument that interfaces with the
dock. If the dock detects a PIN in the
instrument that matches its own PIN, it will
process the instrument. If the dock fails to
detect a PIN, or recognizes a non-
matching PIN, it will not process the
instrument.
To use the passcode feature a PIN must
be entered into the MultiPro IQ Express
Software and then uploaded to every
instrument that will interface with the dock.
Each dock must then be programmed with
the same PIN and set to check for the
code in the instruments.
1. Verify that the MultiPro instruments
have firmware version 5.0 or higher.
Instrument firmware version is shown
in the start up screens for the
instrument.
If needed, new MultiPro Firmware can
be downloaded at no charge from the
Biodownload's website at
http://www.biodownloads.com.
2. Verify that the MultiPro IQ Express
Dock contains firmware version 5.0 or
higher. New dock firmware can also
be downloaded at no charge from the
Biodownload's website at
http://www.biodownloads.com.
3. Select the dock that will be used to
write the PIN (Personal Identification
Number) to the instrument. Connect
the dock to the PC either by USB port
or by Ethernet port. If using an
Ethernet connection, the Ethernet
setting must be configured to "Live". If
more than one dock will be used to
update the instruments, perform steps
4-10 for each dock.
4. Open the IQ Express software.
5. Click "Advanced" in the row of buttons
on the lower right. The software will
automatically show the Security Login
screen and prompt you for your user
name and password. If the
"Advanced" button is not shown, click
on "Change Config" first to access it.
6. Enter a valid User ID and Password.
The Access PIN settings will be
shown:
18