1. Overview
Main unit:
A
B
Automatic Blood
F
C
Pressure Monitor
E
D
Model HEM-7203
A. Display
Instruction Manual
B. START/STOP button
C. Memory button
5327868-4B
D. Up/Down (
) buttons
E. Date/Time setting (
) button
Introduction
Thank you for purchasing the OMRON HEM-7203 Automatic Blood Pressure
Monitor.
Arm cuff:
The OMRON HEM-7203 is a compact, fully automatic blood pressure monitor,
I
J
K
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the devices uses its advanced
"IntelliSense" technology.
Intended Use
L
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
Storage case:
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT
YOUR DOCTOR.
Display:
Important Safety Information
M
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
N
or arteriosclerosis.
O
Please read this section carefully before using the unit.
Warning:
P
• Indicates a potentially hazardous situation which, if not avoided, could result
Q
in death or serious injury.
R
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
S
self-treatment are dangerous.
M. Systolic blood pressure
• People with severe blood flow problems, or blood disorders, should consult a
N. Diastolic blood pressure
doctor before using the unit, as cuff inflation can cause internal bleeding.
(AC Adapter Usage)
O Heartbeat symbol
• Never plug in or unplug the power cord from the electric outlet with wet
(Flashes during measurement)
P. Memory symbol
hands.
(Battery Usage)
Q. Deflation symbol
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
R. Average value symbol
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
2. Preparation
minor or moderate injury to the user or patient or damage to the equipment or
other property.
2.1 Installing/Replacing the Batteries
(General Usage)
1.
• Do not leave the unit unattended with infants or persons who cannot express
Remove the battery cover.
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
2.
Insert four "AA" batteries
• Do not disassemble the unit or arm cuff.
as indicated in the battery
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
compartment and then
in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body
replace the battery cover.
when taking measurements at night. This could result in injury when the air
pressure in the air tube is increased.
Notes:
• If the battery low symbol (
) appears on the display, turn the unit off
• Do not leave the cuff wrapped on the arm if taking measurements during the
night. This could result in injury.
then replace all batteries at the same time.
• Do not inflate the arm cuff over 299 mmHg.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
Disposal of used batteries should be carried out in accordance with the
• Do not operate the unit in a moving vehicle (car, airplane).
national regulations for the disposal of batteries.
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it
can cause internal bleeding.
(AC Adapter Usage)
2.2 Setting the Date and Time
• Use only the original AC adapter (optional) designed for this unit. Use of
unsupported adapters may damage and/or may be hazardous to the unit.
1.
Press the
button.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
2.
Set the monitor to the correct date and time before
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
taking a measurement for the first time.
the power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
1)
plenty of clean water.
• Use only four "AA" alkaline or manganese batteries with this unit. Do not use
other types of batteries.
2)
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
Year
Month Day
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking,
3.
Press the START/STOP button to store the setting.
exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
Notes:
• When removing the air tube, pull on the air plug at the connection with the
• If the batteries have been removed for 30 seconds or more, the date
main unit not the tube itself.
and time setting will need to be reset.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• If the date and time are not set, "-:--" appears during or after
• Do not wash the arm cuff or immerse it in water.
measurement.
• Read and follow the "Important information regarding Electro Magnetic
Compatibility (EMC)" in the Technical Data Section.
• Read and follow the "Correct Disposal of This Product" in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
3. Using the Unit
G
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm.
Do not place the cuff over thick clothes.
H
1.
Insert the air plug into the air jack.
F. Air Jack
G. Battery compartment
H. AC adapter jack (for optional AC
adapter)
2.
Put your arm through the cuff loop.
I. Arm cuff (Medium cuff: arm
circumference 22-32cm)
J. Marker (arrow under tube)
K. Air tube
L. Air plug
3.
Position the arm correctly.
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
T
Marker (arrow under tube) is centred on the middle of your inner arm.
Close the fabric fastener FIRMLY.
U
1 to 2 cm
V
S. Date/Time display
Notes:
T. Blood pressure level indicator
• When you take a measurement on the right
U. Pulse display and Memory
arm, air tube will be at the side of your
number
elbow. Be careful not to rest your arm on
V. Battery low symbol
the air tube.
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. Omron recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your doctor which arm to use for your
measurement.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. No eating,
smoking or exercising 30 minutes before
taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your
heart.
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button to turn the
unit off and to release the air in the arm cuff.
• Remain still while taking a measurement.
1.
Press the START/STOP button.
The cuff will start to inflate automatically.
Push
or
buttons to
START
INFLATE
DEFLATE
END
COMPLETE
change.
Push the
button to confirm.
Next setting appears.
Hour Minute
If your systolic pressure is more than
220 mmHg
After the cuff starts to inflate, press and hold the
START/STOP button until the monitor inflates 30 to 40 mmHg
higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Note: Wait 2-3 minutes before taking another blood pressure
4. Troubleshooting and Maintenance
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
4.1 The Icons and Error Messages
measurement.
Error Display
Cause
Warning:
Self-diagnosis of measured results and treatment are dangerous.
Blink
The batteries are low.
Please follow the instructions of your doctor.
The batteries are
Lit
2.
exhausted.
Remove the arm cuff.
3.
Press the START/STOP button to turn the monitor off.
Cuff is under inflated.
The monitor automatically stores the measurement in its
memory.
It will automatically turn off after five minutes.
Movement during
measurement.
Important:
• Recent research suggests that the following
values can be used as a guide to high blood
Air plug disconnected.
pressure for measurements taken at home.
Arm cuff not applied
correctly.
Systolic Blood Pressure
Above 135 mmHg
Clothing is interfering with
the arm cuff.
Diastolic Blood Pressure
Above 85 mmHg
This criteria is for home blood pressure measurement.
Air is leaking from the arm
cuff.
3.4 Using the Memory Function
The arm cuff was inflated
above 299 mmHg when
The monitor automatically stores the result up to 30 sets. It can also
inflating the cuff manually.
calculate an average reading based on the measurements from the last
three readings taken within 10 minutes. If there are only two readings in
memory for that period, the average will be based on two readings. If
there is one reading in memory for that period, the average will be
Device error.
based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time,
"-:--" is displayed instead of the date and time.
Average Value
Previous Reading
4.2 Troubleshooting
Alternating
date/time display
Problem
Cause
Arm cuff not applied
correctly.
Old
Recent
The reading is
Movement or talking
extremely low (or
during measurement.
View the Average
View the Previous
View the Readings
high).
Value
Reading
Stored in Memory
• If the previous
• The Memory number
reading was taken
appears for a second
Clothing is interfering
without setting the
before the pulse rate
with the arm cuff.
date and time, the
is displayed. The
average value is not
newest set is
calculated.
numbered "1".
The air tube is not
securely connected
Arm cuff pressure
into the main unit.
To delete all the values stored in memory
does not rise.
Air is leaking from the
When the memory symbol (
) appears, first press the Memory
arm cuff.
button. Then while holding it down, press the START/STOP button
simultaneously for about 2-3 seconds.
Arm cuff deflates too
The arm cuff is loose.
soon.
Cannot measure or
The arm cuff has not
readings are too low
been inflated
or too high.
sufficiently.
Note: You cannot partially delete the stored readings.
The batteries are
empty.
Nothing happens
when you press the
The batteries have
buttons.
been inserted
incorrectly.
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
Other problems.
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
4.3 Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme
Remedy
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
You should replace them
with new ones ahead of
• Do not inflate the arm cuff over 299 mmHg.
time.
• Do not disassemble the unit.
Refer to section 2.1.
• Do not subject the unit to strong shocks or vibrations
You should replace them
(for example, dropping the unit on the floor).
with new ones at once.
• Do not use volatile liquids to clean the main unit. The
Refer to section 2.1.
unit should be cleaned with a soft, dry cloth.
Carefully read and repeat
• Use a soft, moistened cloth and soap to clean the arm
the steps listed under
cuff.
section 3.3.
• Do not wash the arm cuff or immerse it in water.
Repeat measurement.
• Do not use petrol, thinners or similar solvents to clean
Remain still and do not talk
the arm cuff.
during measurement.
Refer to section 3.2.
Insert the air plug securely.
Refer to section 3.1.
Apply the arm cuff correctly.
Refer to section 3.1.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Replace cuff with new one.
Refer to Chapter 5.
Do not inflate the arm cuff
• Do not carry out repairs of any kind yourself. If a defect
above 299 mmHg.
Refer to section 3.3.
occurs, consult your local authorised OMRON
distributor or dealer.
Contact your OMRON retail
Calibration and Service
outlet or distributor.
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your local authorised
OMRON distributor or dealer.
Remedy
4.4 Storage
Apply the arm cuff
Keep the unit in its storage case when not in use.
correctly.
Refer to section 3.1.
1.
Unplug the air tube from the air connector.
Remain still and do not
talk during
2.
Gently fold the air tube into the arm cuff.
measurement.
Refer to section 3.2.
Note:
Remove any clothing
Do not bend the air tube excessively.
interfering with the arm
cuff.
Refer to section 3.2.
Make sure that the air
3.
tube is connected
Place the arm cuff and main unit in the
securely.
storage case.
Refer to section 3.1.
Replace the arm cuff
with a new one.
Do not store the unit in the following situations:
• If the unit is wet.
Refer to Chapter 5.
• Locations exposed to extreme temperatures,
Apply the cuff correctly
humidity, direct sunlight, dust or corrosive
so that it is firmly
vapours.
wrapped around the arm.
• Locations exposed to vibrations, shocks or where it will be at an
Refer to section 3.1.
angle.
Inflate the cuff so that it
is 30 to 40 mmHg above
your previous
measurement result.
5. Optional Parts
Refer to section 3.3.
Replace the batteries
Medium Arm Cuff
Large Arm Cuff
AC Adapter "R"
Arm circumference
Arm circumference
with new ones.
Refer to section 2.1.
22 - 32 cm
32 - 42 cm
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
Using the Optional AC Adapter
1.
Insert the AC adapter plug
into the AC adapter jack on
the rear side of the main
unit.
2.
Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
main unit.
HEM-7203-AP_A_M03_100518.pdf
6. Technical Data
Product Description
Automatic Blood Pressure Monitor
Model
HEM-7203
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180/min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Memory
30 Measurements with date and time
Rating
DC6V
4W
Power Source
4 "AA" batteries 1.5V or AC adapter
(optional)
Battery Life
Capacity of new alkaline batteries is approx.
1500 measurements
Applied Part
= Type B
Protection Against
Internally powered ME equipment
Electric Shock
(When using only the batteries)
= Class II ME equipment
(Optional AC adapter)
Operating
+10°C to +40°C / Maximum: 30 to 90% RH
temperature/ Humidity
Storage temperature/
-20°C to +60°C / Maximum: 10 to 95% RH
Humidity/ Air pressure
/ 700-1060 hPa
Console Weight
Approximately 340g without batteries
Cuff Weight
Approximately 120g
Outer Dimensions
Approximately 123 (w) mm × 85 (h) mm × 141(l) mm
Cuff Dimensions
Approximately 146 mm × 446 mm (Medium cuff: arm
circumference 22 to 32 cm)
Cuff Material
Nylon and polyester
Package Content
Main unit, cuff, instruction manual, storage case,
battery set
Notes:
• Subject to technical modification without prior notice.
• Disposal of this product and used batteries should be carried out in
accordance with the national regulations for the disposal of electronic
products.
Please read this instruction manual carefully before using the device.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Noninvasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical
device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct
operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any hazardous substances.
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 J
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
As
ia Pacific HQ
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg
OMRON (DALIAN) CO., LTD.
Production Facility
Dalian, China
This OMRON product is manufactured under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan for assembly.

Advertisement

loading

Summary of Contents for Omron HEM-7203

  • Page 1 Applied Part = Type B the arm cuff. measurement. during measurement. The OMRON HEM-7203 is a compact, fully automatic blood pressure monitor, Important: operating on the oscillometric principle. It measures your blood pressure and Refer to section 3.2. Protection Against Internally powered ME equipment I.
  • Page 2 ) 按鈕 血壓計會自動將測量值儲存在其記憶中。 • 請勿使用汽油、稀釋劑或類似的溶劑清潔臂帶。 電源 4 節 “AA” 電池 1.5V 或交流/直流適配器 日期/時間設定( ) 按鈕 電池壽命 使用新的堿性電池大約可以進行 1500 次測量 將手臂穿過臂帶回環。 五分鐘後,血壓計會自動關閉。 感謝您購買 OMRON HEM-7203 自動血壓計。 重複測量。測量期間,保 臂帶: 測量過程中移動身體。 持靜坐並且不要說話。 應用的部件 = B 型 OMRON HEM-7203 是一款小巧精致、全自動血壓計,根據示波法原 請參閱第 節。 重要事項: 理操作。可方便快捷地測量您的血壓和脈搏速率。本設備利用先進的...