christina
April 5, 2025

I changed the batteries and now I can’t get the monitor to let me change the date or set the date

1 comments:
Mr. Anderson
April 5, 2025

To set the date on the Omron BP785N blood pressure monitor after changing the batteries:

1. Press the Date/Time setting button. The year will flash on the display.
2. Use the up or down button to change the year.
3. Press the button to confirm. The month will flash.
4. Repeat the same steps to set the month, day, hour, and minute.
5. Press the START/STOP button to turn the monitor off after setting is complete.

This answer is automatically generated

Rick Schindler
March 12, 2025

My Omron BP786N always reads higher than doctor's office monitor. How do I reset it?

1 comments:
Mr. Anderson
March 12, 2025

To reset the Omron BP785N blood pressure monitor to take a single measurement instead of three, you need to adjust the TruRead Mode. Press the clock-like symbol to access the TruRead settings. Use the arrows to turn the mode off. Once set, press the Start/Stop button to save and exit.

This answer is automatically generated

robert pendergrass
March 9, 2025

update date

1 comments:
Mr. Anderson
May 14, 2025

The update date for the Omron BP785N is April 28, 2025, based on the review mentioning a verified purchase on that date.

This answer is automatically generated

robert pendergrass
March 9, 2025

change date

1 comments:
Mr. Anderson
March 9, 2025

To change the date on the Omron BP785N blood pressure monitor:

1. Press the button to enter the date setting mode.
2. The year will start flashing. Use the arrow buttons to adjust the year.
3. Press the button again to move to the month setting. Use the arrows to change the month.
4. Press the button again to move to the day setting and adjust it with the arrows.
5. Continue pressing the button to cycle through and confirm the settings.

Once finished, the date and time will be set, and the monitor will be ready for use.

This answer is automatically generated

D
February 21, 2025

My machine takes a measurement but when it is supposed to show blood pressure it starts over

1 comments:
Mr. Anderson
February 21, 2025

The provided context does not contain specific troubleshooting steps for the issue where the Omron BP785N restarts instead of displaying the blood pressure reading. However, general troubleshooting steps for electronic blood pressure monitors include:

1. Check the Power Source – Ensure the batteries are properly installed or the AC adapter is securely connected.
2. Reset the Device – Remove and reinsert the batteries or unplug and reconnect the power adapter.
3. Proper Usage – Make sure the cuff is correctly positioned and you are following the correct measurement procedure.
4. Avoid Interference – Ensure there are no electronic devices nearby that may cause interference.
5. Consult the Manual – Refer to the instruction manual for troubleshooting guidance.
6. Contact Support – If the issue persists, consider reaching out to Omron customer support.

If the issue continues, professional inspection may be needed.

This answer is automatically generated

Craig
February 13, 2025

Blood pressure reading high compared to Dr office reading

1 comments:
Mr. Anderson
May 14, 2025

The context does not specifically explain why the Omron BP785N blood pressure reading might be higher than the reading at the doctor's office. However, it mentions that factors like stress, improper resting time, talking, or incorrect arm positioning during measurement can cause inaccurate readings. It also advises following specific steps before measuring, such as resting for at least 15 minutes and avoiding exercise, caffeine, smoking, or eating 30 minutes prior. Differences in measurement techniques between home and clinic settings could also contribute to variations.

This answer is automatically generated

connie
February 2, 2025

how to change mode from average of three bp to one time reading

1 comments:
Mr. Anderson
February 10, 2025

To change the mode from averaging three blood pressure readings to a one-time reading on the Omron BP785N, follow these steps:

1. Turn on the device but do not start a measurement.
2. Press the button that controls the reading mode.
3. Look for the option that enables or disables the three-reading average mode.
4. Use the arrow button to switch it off.
5. Confirm the setting by pressing the button again.

This will set the device to take a single blood pressure reading instead of averaging three.

This answer is automatically generated

Henrietta Ransom
January 22, 2025

Machine is no longer tracking info on my phone

1 comments:
Mr. Anderson
February 10, 2025

The provided context does not mention any phone connectivity features for the Omron BP785N. If your device is not tracking information on your phone, check if it supports Bluetooth or other connection methods. If it does, ensure that Bluetooth is enabled, the device is paired correctly, and any required app is installed and functioning. If the issue persists, refer to the instruction manual or contact customer support.

This answer is automatically generated

DWAYNE CECIL REDGER
July 29, 2025

HOW DO YOU SET THE BLOOD PRESSER MACHINE TO TAKE THE BLOOD PRESSER ONE TIME NOT THREE TIMES BEFORIT DISPLAYS THE INFORMATION

Shirley Blair
June 19, 2025

Can I copy it BP to paper monthly

User image 6853fe4b46c38

Subscribe to Our Youtube Channel

Summary of Contents for Omron BP785N

  • Page 1 English Español 10 series Blood Pressure Monitor Model BP785N Instruction Manual 2298861-6C...
  • Page 2 Product includes: El producto incluye: AC Adapter Adaptador de CA ComFit™ Cuff Brazalete ComFit™ Monitor Monitor Quick Start Guide Guía resumida Instruction Manual Manual de Instrucciones...
  • Page 3: Table Of Contents

    Contents ® Thank you for purchasing the OMRON BP785N Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
  • Page 4: Important Safety Information

    Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. The monitor is not intended to be a diagnostic device.
  • Page 5 Important Safety Information Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. (General Usage) Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: •If you have had a mastectomy.
  • Page 6 Wipe the dust off from the power plug. Unplug monitor when not in use. Disconnect the power plug before cleaning. Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device. (Battery Usage) Do not insert the batteries with their polarities incorrectly aligned.
  • Page 7: Know Your Device

    1. Know Your Device Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. Monitor Display Air jack Memory symbol Blood pressure color indicator USER ID symbol Date/Time setting button Average value symbol...
  • Page 8: Display Symbols

    1. Know Your Device 1.1 Display Symbols Irregular Heartbeat Symbol ( Normal Heartbeat When the monitor detects an irregular rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol will appear on the display with Blood pressure the measurement values.
  • Page 9: Before Taking A Measurement

    1. Know Your Device Blood Pressure Color Indicator If your Systolic Blood Pressure is 135 mmHg or above and/or the Diastolic Blood Pressure is 85 mmHg or above, the blood pressure color indicator will light in “orange” when the measurement result is displayed. If the results are within the standard range, the blood pressure color indicator will light in “green”.
  • Page 10: Preparation

    2. Preparation 2.1 Battery Installation Remove the battery cover. Insert 4 “AA” batteries as indicated in the battery compartment. Replace the battery cover. Notes: • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time.
  • Page 11: Using The Ac Adapter

    AC Adapter is disconnected. The measurement results will not be deleted. Use only OMRON authorized parts and accessories. Parts and accessories not approved for use with the device may damage the device.
  • Page 12 2. Preparation Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.
  • Page 13: Setting The Trureadtm Mode

    2. Preparation 2.3 Setting the TruRead Mode The TruRead Mode takes 3 consecutive measurements. The monitor will inflate, take a measurement, and deflate - 3 times, separated by a short interval between each measurement. The TruRead Mode is set “oFF” by default. Select your USER ID (1 or 2).
  • Page 14: Setting The Date And Time

    2. Preparation Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. 2.4 Setting the Date and Time Set the monitor to the correct date and time before taking a measurement for the first time. Press the Date/Time setting ( ) button.
  • Page 15: Using The Device

    • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your...
  • Page 16: How To Sit Correctly

    3. Using the Device 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. •...
  • Page 17: Taking A Measurement

    3. Using the Device 3.3 Taking a Measurement Notes: • To stop the measurement, press the START/STOP button once to deflate the arm cuff. • Remain still and quiet while taking a measurement. The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2.
  • Page 18: Using The Truread Tm Mode

    3. Using the Device Remove the arm cuff. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
  • Page 19 3. Using the Device The first measurement is complete, waiting to start the next measurement. The interval will appear on the display. The second measurement is complete, waiting to start the next measurement. The interval will appear on the display. The third measurement is complete, the average for the 3 measurements appears on the display.
  • Page 20: Using The Guest Mode

    3. Using the Device Using the Guest Mode The monitor stores measurement values for 2 users in the memory. The guest mode can be used to take a single measurement for another user. No measurement values are stored in the memory, and the TruRead mode is not available when the guest mode is selected.
  • Page 21: Using The Memory Function

    3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 100 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes.
  • Page 22 3. Using the Device Press the button to view the values stored in the memory. : To view the older values : To view the more recent values Note: The value with the TruRead symbol is the average for the consecutive 3 measurements. To display the individual measurement values, press the button while the average is displayed.
  • Page 23: Deleting All Values Stored In Memory

    3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. Select your USER ID (1 or 2). Press the Memory button, while the memory symbol ( ) appears.
  • Page 24: Error Messages And Troubleshooting

    4. Error Messages and Troubleshooting 4.1 Error Messages Display Cause Solution Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Irregular heartbeats are detected. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Carefully read and repeat the steps in section Movement during measurement.
  • Page 25 4. Error Messages and Troubleshooting Display Cause Solution Insert the air plug securely. Air plug disconnected. Refer to section 3.1. Apply the arm cuff tighter. Arm cuff is applied too loosely. Refer to section 3.1. Replace the arm cuff with a new one. Air is leaking from the arm cuff.
  • Page 26: Troubleshooting

    4. Error Messages and Troubleshooting 4.2 Troubleshooting Problem Cause and Solution Replace all batteries with new ones. No power. Check the battery installation for proper placement of the battery polarities. No display appears on the monitor. Refer to section 2.1. Blood pressure varies constantly.
  • Page 27: Maintenance And Storage

    5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please observe the following: • Store the device and the components in a clean, safe location. • Do not use any abrasive or volatile cleaners. • Do not wash the device and any components or immerse them in water. •...
  • Page 28: Storage

    5. Maintenance and Storage 5.2 Storage Unplug the air plug from the air jack. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. Do not store the device in the following situations: •...
  • Page 29: Optional Medical Accessories

    5. Maintenance and Storage 5.3 Optional Medical Accessories Arm Cuff AC Adapter Arm circumference 9” - 17” (22 - 42 cm) HEM-ADPTW5 CFX-WR17 (Model: HEM-FL31)
  • Page 30: Specifications

    6. Specifications Model BP785N HEM-7321-Z Display LCD digital display Measurement range Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min. Accuracy Pressure: ±3 mmHg or 2% of reading Pulse: ± 5% of display reading Inflation Fuzzy-logic controlled by electric pump...
  • Page 31: Fcc Statement

    7. FCC Statement FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
  • Page 32: Limited Warranty

    8. Limited Warranty Your BP785N Automatic Blood Pressure Monitor, excluding the arm cuff, is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor.
  • Page 33: Guidance And Manufacturer's Declaration

    • The MEDICAL ELECTRICAL EQUIPMENT BP785N including AC-adapter needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this documentations. • The Essential Performance of the BP785N including AC-adapter is to measure a blood pressure and a pulse rate and using the memory function.
  • Page 34 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP785N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP785N including AC-adapter should assure that it is used in such environment.
  • Page 35 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP785N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP785N including AC-adapter should assure that it is used in such environment.
  • Page 36 OMRON BP785N including AC-adapter OMRON BP785N including AC-adapter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this OMRON BP785N including AC-adapter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the OMRON BP785N including AC-adapter as recommended below, according to the maximum output power of the communications equipment.
  • Page 40 Manufactured for : OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN Distributed by : OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A. www.omronhealthcare.com © 2014 OMRON HEALTHCARE, INC. Made in Vietnam Fabricado para : OMRON HEALTHCARE Co., Ltd.

This manual is also suitable for:

Bp786n10 series

Table of Contents

Save PDF